Keypad Display Menu Tree of LevelPRO Gauges
MAIN MENU
APPS
SysStatus
DETECTORS
SYS CNTRL
PHYSCL I/O
COM SETUP
I/O CAL
PASSWORDS
COMMANDS
WrmCldStr
Mod/Flt Alm
IO CALIBRATION
mA#1
Cur Out A
mA#2
Cur Out B
Vdc#1
Cur Out C
Vdc#2
Det Cur In
RTD INPUT
DETECTOR SELECT
SELECT DETECTOR
NUMBER
1
LIST OF DETECTORS
PRESENT
NONE NONE NONE DET1
DetectorCurrentInput
VIEW CAL POINTS
BEGIN CALIBRATION
DetectorCurrentInput
CALIBRATION POINTS
3PointCal
mA Value¦ RawValue
Min 4.000000¦
4.000
Mid 12.00000¦
12.000
Max 20.00000¦
20.000
BACK
NEXT
BACK
NEXT
BACK
CHANGE
NEXT
BACK
NEXT
BACK
NEXT
DetectorCurrentInput
VIEW CAL POINTS
BEGIN CALIBRATION
CALIBRATION POINTS
ENTER
CAL POINT:
3PointCal
OR
PRESS NEXT TO START
CALIBRATION
ENTER NEW min VALUE
4.000000
RAW min VALUE
0.000
PRESS FREEZE WHEN OUT
PUT STABLE ENTER VALU
HIT SUBMIT THEN NEXT
BACK
NEXT
BACK
CHANGE
NEXT
BACK
FREEZE
NEXT
ENTER NEW mid VALUE
12.00000
RAW mid VALUE
0.000
PRESS FREEZE WHEN OUT
PUT STABLE ENTER VALU
HIT SUBMIT THEN NEXT
ENTER NEW max VALUE
20.00000
RAW max VALUE
0.000
PRESS FREEZE WHEN OUT
PUT STABLE ENTER VALU
HIT SUBMIT THEN NEXT
BACK
FREEZE
NEXT
BACK
FREEZE
NEXT
DetectorCurrentInput
ACCEPT CAL?
BACK
YES
NO
IO CALIBRATION
mA#1
Cur Out A
mA#2
Cur Out B
Vdc#1
Cur Out C
Vdc#2
Det Cur In
RTD INPUT
DETECTOR SELECT
SELECT DETECTOR
NUMBER
1
LIST OF DETECTORS
PRESENT
NONE NONE NONE DET1
RTD INPUT
VIEW CAL POINTS
BEGIN CALIBRATION
RTD INPUT
CALIBRATION POINTS
3PointCal
EU Value¦ RawValue
Min 0.000000¦
0.0
Mid 50.00000¦
50.0
Max 100.0000¦
100.0
BACK
NEXT
BACK
CHANGE
NEXT
BACK
NEXT
BACK
NEXT
RTD INPUT
VIEW CAL POINTS
BEGIN CALIBRATION
CALIBRATION POINTS
ENTER
CAL POINT:
3PointCal
OR
PRESS NEXT TO START
CALIBRATION
ENTER NEW min VALUE
0.000000
RAW min VALUE
-253.0
PRESS FREEZE WHEN OUT
PUT STABLE ENTER VALU
HIT SUBMIT THEN NEXT
BACK
NEXT
BACK
CHANGE
NEXT
BACK
FREEZE
NEXT
ENTER NEW mid VALUE
50.00000
RAW mid VALUE
-253.0
PRESS FREEZE WHEN OUT
PUT STABLE ENTER VALU
HIT SUBMIT THEN NEXT
ENTER NEW max VALUE
100.0000
RAW max VALUE
-253.0
PRESS FREEZE WHEN OUT
PUT STABLE ENTER VALU
HIT SUBMIT THEN NEXT
BACK
FREEZE
NEXT
BACK
FREEZE
NEXT
RTD INPUT
ACCEPT CAL?
BACK
YES
NO
Figure A-11.
Input/Output Calibration; Detector Current & RTD Inputs
Thermo Scientific
LevelPRO User Manual
A-15
Содержание LevelPRO
Страница 2: ......
Страница 4: ......
Страница 6: ...This page intentionally left blank ...
Страница 8: ...This page intentionally left blank ...
Страница 16: ......
Страница 102: ......
Страница 127: ......