Flash Application Firmware
Thermo Scientific
LevelPRO User Manual
B-3
5.
From the Functions dropdown on the menu bar, select Flash Application
Firmware to bring up the following screen. This screen can also be accessed by
clicking the second icon button, circled in the figure below.
Figure B-3.
Flash Application Screen
6.
Select the firmware to flash by clicking either Main Board Firmware S19 or IBP
Board Firmware Hex.
Note:
If programming the IBP (Integrated Backplane) board, the Main Board
must be removed from the integrated unit or the RS485 wire must be
disconnected from the remote unit.
7.
Choose the correct .S19 or .hex file and click the Open button.