Advanced Maintenance
Replacing the Covers of the ISQ 7000 Mass Spectrometer
276
ISQ 7000 Hardware Manual
Thermo Scientific
Top Cover Panel
To replace the top cover panel of the ISQ 7000 instrument
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Use a T20 Torxhead screwdriver to loosen the top cover panel screw from the interior
front panel. Then slide the top cover panel toward the back of the instrument and lift it
off.
3. Replace the top cover panel and tighten the screw holding it in place.
Restart your system by following the instructions in
.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge or the column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.
Figure 290.
Replacing the Top Cover Panel of the ISQ 7000 Instrument
Top Cover Panel Screw
Top Cover Panel