5
Operation
Starting operation
Thermo Scientific
HERA
CELL VIOS
160i
/
250i
5-5
Loading the device:
18. Load the workspace with cultures.
Note
Duration of the auto-start routine:
When the device is cold and when the ambient temperature is low, the auto-start routine may take
up to 10 hours.
Loading:
To ensure sufficient air circulation and even heating of the samples, the loading surface within the
workspace should be used up to 70% max. Bulky objects in the workspace that dissipate heat may
impair heat distribution. Bulky objects in the use able space that dissipate heat may impair heat
distribution.