![Thermo Scientific HERACELL VIOS 160i Скачать руководство пользователя страница 171](http://html1.mh-extra.com/html/thermo-scientific/heracell-vios-160i/heracell-vios-160i_user-manual_1104556171.webp)
12
Annex: Data communication
Overview of general parameters (addresses 0xxx)
Thermo Scientific
HERA
CELL VIOS
160i
/
250i
12-5
Description of the response elements:
aaaa
parameter address,
bb
amount of payload (always 02)
cc
Checksum: Inverted XOR of all bytes without
checksum and <CR>
XX = 2 bytes error message (see table below)
Example of an unknown command:
Query:
?:0005:00::cc<CR>
Response
!:0005:02:?1:cc<CR>
Significance of the two bytes in the error message:
Overview of general parameters (addresses 0xxx)
General parameters are system values such as date, time, and motherboard version number.
Reading general parameters
Overview of incubator parameters (addresses 2xxx)
Incubator parameters are divided into:
• the basic parameters of the three control loops temperature, CO
2
and O
2
,
• the internal function parameters of operating functions and data logging.
Failure message
Description
?0
Error in telegram structure or check sum
?1
Unknown error or unknown parameter
?2
Internal memory error
?3
Data error (value not within set limits)
Address
Description
Remarks
0001
Mainboard version number
8 places
0010
Date and time display
[hours:minutes:seconds];[day:mon
th:year]
17 bytes / decimal value
with the format xx:xx:xx;xx:xx:xx
0011
Date
[day:month:year]
8 bytes / decimal value
with the format xx:xx:xx
0012
Time
[hours:minutes:seconds]
8 bytes / decimal value
with the format xx:xx:xx
Содержание HERACELL VIOS 160i
Страница 8: ...Table of Contents 6 HERACELL VIOS 160i 250i Thermo Scientific ...
Страница 12: ...Figures 4 HERACELL VIOS 160i 250i Thermo Scientific ...
Страница 76: ...5 Operation Starting operation 5 6 HERACELL VIOS 160i 250i Thermo Scientific ...
Страница 128: ...6 Handling and control Measures after power outage 6 52 HERACELL VIOS 160i 250i Thermo Scientific ...
Страница 154: ...9 Maintenance Replacing the Door Seal 9 12 HERACELL VIOS 160i 250i Thermo Scientific ...
Страница 166: ...11 Specifications HERACELL VIOS 250i 11 10 HERACELL VIOS 160i 250i Thermo Scientific ...
Страница 196: ...13 Device log 13 2 HERACELL VIOS 160i 250i Thermo Scientific ...