![Thermo Scientific HERACELL VIOS 160i Скачать руководство пользователя страница 124](http://html1.mh-extra.com/html/thermo-scientific/heracell-vios-160i/heracell-vios-160i_user-manual_1104556124.webp)
6
Handling and control
Measures after power outage
6-48
HERA
CELL VIOS
160i
/
250i
Thermo Scientific
Troubleshooting
The error tables indicate source of error, cause of error, and possible corrections.
For any communication with the Technical Service of Thermo Fisher Scientific, please have the device
data ready.
Control
loop
Failure message
Cause
Remedy
Alarm
relay
Audible
alarm
Log
System
Device door open too
long
Device door has been
open for longer than 10
minutes
Close device door
X
X
X
Error: Display
Display not
communicating with
the main board
*1)
Reset unit. If the
error occurs
repeatedly, call
service.
X
X
X
Error: EEPROM main
board
EEPROM on main
board is defective
Reset unit. If the error
occurs repeatedly, call
service.
X
X
X
Error: data logger
Error while writing to
the memory of the data
logger. Incubator is still
operative.
Reset unit. If the error
occurs repeatedly, call
service.
Error: steri-run
Error in Steri-Run
routine
Reset unit. If the error
occurs repeatedly, call
service.
X
X
X
Power down during
steri-run
Power failure during
Steri-Run routine
Restart unit and
re-launch Steri-Run.
X
X
X
Error: auto-start
Error in auto-start
routine
Re-launch auto-start.
If the error occurs
repeatedly, call ser-
vice.
X
X
X
Error: ADC
Measurement of
reference resistance out
of tolerance
Reset unit. If the error
occurs repeatedly, call
service.
x
x
x
Error: Fan
Actual value of the fan
is out of tolerance.
Reset unit. If the error
occurs repeatedly, call
service.
X
X
X
IR sensor changed
New serial number
detected
Acknowledge the
alarm
X
X
X
Содержание HERACELL VIOS 160i
Страница 8: ...Table of Contents 6 HERACELL VIOS 160i 250i Thermo Scientific ...
Страница 12: ...Figures 4 HERACELL VIOS 160i 250i Thermo Scientific ...
Страница 76: ...5 Operation Starting operation 5 6 HERACELL VIOS 160i 250i Thermo Scientific ...
Страница 128: ...6 Handling and control Measures after power outage 6 52 HERACELL VIOS 160i 250i Thermo Scientific ...
Страница 154: ...9 Maintenance Replacing the Door Seal 9 12 HERACELL VIOS 160i 250i Thermo Scientific ...
Страница 166: ...11 Specifications HERACELL VIOS 250i 11 10 HERACELL VIOS 160i 250i Thermo Scientific ...