6
Handling and control
Measures after power outage
Thermo Scientific
HERA
CELL VIOS
160i
/
250i
6-49
Tem-
perature
Sensor breakage
Measured value exceeds
accepted limit
Request service.
Dry the heaters.
X
X
X
Actual value above
Actual value > Set value
+ 1 °C
*2) *4)
Do
not exceed
permissible ambient
temperature / call
field service.
X
X
X
Actual value low
Actual value > Set value
+ 1 °C
*3) *4)
If the error does not
reset automatically,
call field service.
X
X
X
Actual value not
plausible
Implausible
temperature signal
Reset unit. If the
error occurs
repeatedly, call
service.
X
X
X
Calibration values too
high/low
Max. adjust value for
temperature exceeded /
not attained
Acknowledge alarm,
enter different target
value.
X
Control
loop
Failure message
Cause
Remedy
Alarm
relay
Audible
alarm
Log
Содержание HERACELL VIOS 160i
Страница 8: ...Table of Contents 6 HERACELL VIOS 160i 250i Thermo Scientific ...
Страница 12: ...Figures 4 HERACELL VIOS 160i 250i Thermo Scientific ...
Страница 76: ...5 Operation Starting operation 5 6 HERACELL VIOS 160i 250i Thermo Scientific ...
Страница 128: ...6 Handling and control Measures after power outage 6 52 HERACELL VIOS 160i 250i Thermo Scientific ...
Страница 154: ...9 Maintenance Replacing the Door Seal 9 12 HERACELL VIOS 160i 250i Thermo Scientific ...
Страница 166: ...11 Specifications HERACELL VIOS 250i 11 10 HERACELL VIOS 160i 250i Thermo Scientific ...
Страница 196: ...13 Device log 13 2 HERACELL VIOS 160i 250i Thermo Scientific ...