7
Cleaning and Care
Clean the rotor and the accessories regularly and in
case of need. This is indicated both for reasons of hy-
giene and to prevent corrosion due to contamination
sticking to the instrument and its accessories.
Clean them with mild agents of pH values ranging from
6 to 8.
Immediately after cleaning, dry the aluminum parts or
put them into a warm-air dryer at a temperature not
exceeding 50°C.
Disinfection
If a capillary tube containing infectious material leaks,
during a centrifugation run, you have to disinfect im-
mediately the rotor.
Infectious material could enter the
centrifuge if spills or tube breakage
occur.
Danger of infection may occur upon
contact!
Take appropriate protective measures
for personnel!
Mind the permissible filling volumes
and loading limits for the tubes!
Broken glass capillaries have sharp
edges and pose a risk of injury.
In case of contamination the operator
has to make sure, that no further per-
sons are jeopardized!
Contaminated parts have to be decon-
taminated immediately.
If required further protective measures
have to be initiated
.
Содержание Haematocrit-Rotor 7500 3473
Страница 1: ...Haematocrit Rotor 7500 3473 ...
Страница 2: ......
Страница 8: ...6 für Ihre Notizen ...
Страница 12: ...10 für Ihre Notizen ...
Страница 18: ...6 for your notes ...
Страница 22: ...10 for your notes ...
Страница 28: ...6 pour vos notes ...
Страница 32: ...10 pour vos notes ...
Страница 38: ...6 Per i vostri appunti ...
Страница 42: ...10 Per i vostri appunti ...
Страница 48: ...6 para sus notas ...
Страница 52: ...10 para sus notas ...
Страница 58: ...6 Voor uw notities ...
Страница 62: ...10 Voor uw notities ...