Installation
Connect the electrical wiring
The main components of the system must be installed on separate circuits. Plug the components into
the two power strips and connect each of the power strips to a separate grounded AC power source
on different circuits:
l
Power strip A:
o
Heated Valve Drawer
o
Inlet Heated Sample Line power
o
Peristaltic pump
l
Power strip B:
o
Spectrometer or analyzer
o
Heated gas cell
o
Computer
Each of the power strips in this setup will draw about 10 A. If they are plugged into the same circuit,
they will overload most standard circuits.
1. Connect the vacuum pump to the back panel of the Heated Valve Drawer.
2. Plug the main power from the Heated Valve Drawer into one of the two power strips (either the
front or back power strip).
3. Plug the peristaltic pump into the same power strip that you plugged the heated valve drawer into.
4. Plug the spectrometer, heated gas cell, and computer into the second power strip.
5. Plug the power strips into two separate grounded AC power sources. The outlets must be on two
different circuits.
30
Modular Gas System User Guide
Thermo Scientific
Содержание FTIR
Страница 1: ...Modular Gas System USERGUIDE 269 346401B July 2022 ...
Страница 15: ... This page intentionally left blank ...
Страница 38: ...using the pressure transducer Thermo Scientific Modular Gas System User Guide 33 Installation ...
Страница 61: ... This page intentionally left blank ...
Страница 79: ... This page intentionally left blank ...
Страница 81: ...Plumbing diagrams Automatic Pressure Control System 76 Modular Gas System User Guide Thermo Scientific ...
Страница 82: ...Eductor Panel Thermo Scientific Modular Gas System User Guide 77 Plumbing diagrams ...
Страница 83: ... This page intentionally left blank ...