The following table lists the specifications for electrical service. Contact our service representative in
your area if you have questions about the requirements.
Requirements
Specifications
Input current
10 A (max.)
Input voltage
230 Vac
Line frequency
50 to 60 Hz
Line disturbances
Sags, surges or other line disturbances must not exceed 10% of input
voltage (even for a half cycle).
Noise
< 2 V (common mode)
< 20 V (normal mode)
Power cords
Use an appropriate power cord for the electrical service. The power cord supplied with the
spectrometer is a 3-wire, grounded power cord, appropriate for use in the country listed as the
shipping destination for the system. The main power supply cord can be replaced only by an
equivalent cable. If the power cord you received is not appropriate for the electrical system in your
location or if the power cord becomes damaged, contact us to order a new power cord.
A 10 A rated power cord must be used for each power cord.
Temperature and humidity
The FTIR Heated Valve Drawer is designed for indoor use at altitudes up to 2,000 M (6,500 ft). The
drawer itself operates reliably at temperatures between 16 °C and 27 °C (60 °F and 80 °F).
Maintain humidity at 20% to 80% non-condensing.
The safe environmental operating temperatures of the gas analysis system at your location may be
restricted by the safe operating parameters for the gas cell(s) used with your analyzer or
spectrometer.
Thermo Scientific
Modular Gas System User Guide
7
Site requirements and safety information
Содержание FTIR
Страница 1: ...Modular Gas System USERGUIDE 269 346401B July 2022 ...
Страница 15: ... This page intentionally left blank ...
Страница 38: ...using the pressure transducer Thermo Scientific Modular Gas System User Guide 33 Installation ...
Страница 61: ... This page intentionally left blank ...
Страница 79: ... This page intentionally left blank ...
Страница 81: ...Plumbing diagrams Automatic Pressure Control System 76 Modular Gas System User Guide Thermo Scientific ...
Страница 82: ...Eductor Panel Thermo Scientific Modular Gas System User Guide 77 Plumbing diagrams ...
Страница 83: ... This page intentionally left blank ...