Important
Read this instruction manual. Failure to read,
understand and follow the instructions in this manual may
result in damage to the unit, injury to operating personnel,
and poor equipment performance.
s
Caution
All internal adjustments and maintenance must be performed by qualified service personnel.
s
This equipment has been designed and tested to conform to CSA1010 safety standards, as applicable to
laboratory instrumentation. This applies only to the EC300 when used as specified in the documentation, in its
intended applications, with Thermo Fisher Scientific approved electrophoresis apparatus only. Usage in any other
manor may not provide similar performance or safety protection.
This equipment is provided with a 3-conductor, grounded AC line cord. The protective earth ground is necessary
for safe operation. Do not use any other AC line cord with this instrument.
The EC300 is a high voltage power supply capable of generating dangerous levels of voltage and current during
operation. Exercise caution when working around and with the electrical connections of this equipment. Always
check electrical connectors, wires, and associated apparatus for any signs of wear or damage before using with
this equipment. Be sure to use only electrophoresis equipment that is suitably rated for the voltage and current
capabilities of the EC300 power supply.
The output of the EC300 power supply is intended for connection to electrically isolated electrophoresis
apparatus only. Use only with electrically isolated electrophoresis apparatus with minimum isolation of 600V. Do
not connect any terminal of the EC300 output to earth ground. This may impair the safety protection provided
by the equipment, or cause equipment damage.
The high voltage output of the EC300 power supply takes some amount of time to decay when unloaded or
lightly loaded. Wait a minimum of 60 seconds after stopping a run before touching the power supply leads.
This equipment has a protective ground leakage current of less than 0.5mA using test methods defined in
CSA1010 and CSA151.
This equipment is for indoor use only.
Material in this manual is for information purposes only. The contents and the product it describes are subject to
change without notice. Thermo Fisher Scientific makes no representations or warranties with respect to this
manual. In no event shall Thermo be held liable for any damages, direct or incidental, arising out of or related to
the use of this manual.