Important

Read this instruction manual. Failure to read,

understand and follow the instructions in this manual may
result in damage to the unit, injury to operating personnel,
and poor equipment performance. 

s

Caution

All internal adjustments and maintenance must be performed by qualified service personnel. 

s

This equipment has been designed and tested to conform to CSA1010 safety standards, as applicable to
laboratory instrumentation. This applies only to the EC300 when used as specified in the documentation, in its
intended applications, with Thermo Fisher Scientific approved electrophoresis apparatus only. Usage in any other
manor may not provide similar performance or safety protection.

This equipment is provided with a 3-conductor, grounded AC line cord. The protective earth ground is necessary
for safe operation. Do not use any other AC line cord with this instrument.

The EC300 is a high voltage power supply capable of generating dangerous levels of voltage and current during
operation. Exercise caution when working around and with the electrical connections of this equipment. Always
check electrical connectors, wires, and associated apparatus for any signs of wear or damage before using with
this equipment. Be sure to use only electrophoresis equipment that is suitably rated for the voltage and current
capabilities of the EC300 power supply.

The output of the EC300 power supply is intended for connection to electrically isolated electrophoresis
apparatus only. Use only with electrically isolated electrophoresis apparatus with minimum isolation of 600V. Do
not connect any terminal of the EC300 output to earth ground. This may impair the safety protection provided
by the equipment, or cause equipment damage.

The high voltage output of the EC300 power supply takes some amount of time to decay when unloaded or
lightly loaded. Wait a minimum of 60 seconds after stopping a run before touching the power supply leads.

This equipment has a protective ground leakage current of less than 0.5mA using test methods defined in
CSA1010 and CSA151.

This equipment is for indoor use only.

Material in this manual is for information purposes only. The contents and the product it describes are subject to
change without notice. Thermo Fisher Scientific makes no representations or warranties with respect to this
manual. In no event shall Thermo be held liable for any damages, direct or incidental, arising out of or related to
the use of this manual.

©2012 Thermo Fisher Scientific. All rights reserved.

Thermo Scientific

ii

EC300XL

Preface

Содержание EC300XL

Страница 1: ...Electrophoresis Power Supply EC300XL Operating and Maintenance Manual 7217581 Rev 0 Visit us online to register your warranty www thermoscientific com warranty ...

Страница 2: ...Thermo Scientific MANUAL NUMBER 72175881 0 5 3 12 Transfer to Marietta was LTM1596X2 4 7 10 ccs REV ECR ECN DATE DESCRIPTION By Preface EC300XL i ...

Страница 3: ...of wear or damage before using with this equipment Be sure to use only electrophoresis equipment that is suitably rated for the voltage and current capabilities of the EC300 power supply The output of the EC300 power supply is intended for connection to electrically isolated electrophoresis apparatus only Use only with electrically isolated electrophoresis apparatus with minimum isolation of 600V ...

Страница 4: ...ment that has been put on the market after 13 August 2005 This product is required to comply with the European Union s Waste Electrical Electronic Equipment WEEE Directive 2002 96 EC It is marked with the WEEE symbol Thermo Fisher Scientific has contracted with one or more recycling disposal companies in each EU Member State European Country and this product should be disposed of or recycled throu...

Страница 5: ...l your needs for spare or replacement parts or provide you with on site service We can also provide you with a quotation on our Extended Warranty for your Thermo Scientific products Whatever Thermo Scientific products you need or use we will be happy to discuss your applications If you are experiencing technical problems working together we will help you locate the problem and chances are correct ...

Страница 6: ...et Up and Operation 2 1 Using the Power Supply 2 2 Constant Current Operation 2 3 Constant Voltage Operations 2 3 Timed Operations 2 4 Automatic Power fail Restart 2 4 Troubleshooting 3 1 Maintenance 4 1 Cleaning 4 1 Replacing a Fuse 4 1 Replacement Parts 4 1 Accessories 4 1 Section 1 Section 2 Section 3 Section 4 ...

Страница 7: ......

Страница 8: ...taneously The EC300 can deliver up to 75W of total output power When operating in constant voltage or constant current mode the power supply automatically limits the other parameter to either the power supply maximum or a lower limit if set by the user If this non constant limit is reached the power supply will automatically switch control modes from constant voltage to constant current or vice ve...

Страница 9: ...your supplier for instructions AC Input Power 100 120VAC 50 60Hz 100VA Max Environmental Operating temperature 0 40 C 0 95 R H non condensing Altitude 2000m Overvoltage category II IEC664 Pollution degree 2 IEC664 DC Output Power 10 300VDC 75W Max 4 400mA 75W Max Ripple 1 Drift 1 after 30 minute warm up Compliance CSA 1010 This equipment has been designed and tested to conform to CSA1010 safety st...

Страница 10: ... to the power supply making sure to match the red positive lead to the red positive jack and the black negative lead to the black negative jack Power the unit on by connecting the AC power cord to the power entry on the rear panel then plugging the 3 prong AC plug into a power source Connect the power supply to a 3 prong grounded AC outlet using the AC cord provided with the unit only Red positive...

Страница 11: ... supply output The power supply output will ramp up to the appropriate setpoint while not allowing the limit parameter to be exceeded If the limit parameter should be exceeded then the power supply will crossover control modes making the limit parameter the new control setpoint The front panel constant mode LED indicator will change to show that a crossover has occurred To stop a run in progress p...

Страница 12: ...tput rating would be exceeded If this value is lower than the limit setpoint entered the EC300 will use this lower limit parameter Press the constant key labeled K until the LED indicating Amps is illuminated Each time the constant key is pressed the EC300 will display the last saved setpoint value for that particular mode Use the UP and DOWN arrow keys to change the setpoint to the desired value ...

Страница 13: ...ey simultaneously press and release the UP arrow key The display will show PF momentarily to indicate power fail restart is active Whenever power fail restart is enabled PF is displayed briefly as the time parameter is selected by pressing the Set key The power fail restart feature can be enabled either before or after changing the time duration to the desired value 0 999 minutes Setting the time ...

Страница 14: ...300 E XX Condition and Possible Remedy 00 Minimum load current detected The output is not connected or the electrophoresis apparatus is not set up properly Check your setup and connections 01 Maximum load current exceeded The output is short circuit or the elec trophoresis apparatus is not set up properly Check your setup and con nections 03 Loss of AC power PF enabled and completed The EC300 dete...

Страница 15: ......

Страница 16: ... EC300 power supply requires no periodic servicing and should provide years of trouble free operation Should you need to replace the fuses proceed as follows Caution The EC300 may use double pole neutral fusing s Turn off AC power by disconnecting the AC line cord Remove the fuse holder assembly with a small flatblade screwdriver Always replace both fuses with appropriate replacement fuses 1 5A 25...

Страница 17: ......

Страница 18: ...nent part beyond the original warranty period The Technical Services Department must give prior approval for return of any components or equipment At Thermo s option all non conforming parts must be returned to Thermo Fisher Scientific postage paid and replacement parts are shipped FOB desti nation THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES WHETHER WRITTEN ORAL OR IMPLIED NO WA...

Страница 19: ... and gaskets are excluded from this warranty Replacement or repair of components parts or equipment under this warranty shall not extend the warranty to either the equipment or to the component part beyond the original warranty period The Technical Services Department must give prior approval for return of any components or equipment At Thermo s option all non conforming parts must be returned to ...

Страница 20: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com ...

Отзывы: