EC300XL 3-1
Thermo Scientific
Section 3
Troubleshooting
The EC300 detects and reports several events and conditions that are
considered errors. The EC300 will stop any run in process and display
"EXX" where XX is one of the below listed errors. When the EC300 is
displaying an error indication, press the blue ON/OFF key to clear the
error and return to Setup mode, or press the Run key to resume the run
(after correcting the cause of the error). Note that loss of AC power during
a run is considered an error condition.
Always stop a run before turning off AC power to the EC300.
E‘XX’
Condition and Possible Remedy
00
Minimum load current detected. The output is not connected, or the
electrophoresis apparatus is not set up properly. Check your setup and
connections.
01
Maximum load current exceeded. The output is short circuit, or the elec-
trophoresis apparatus is not set up properly. Check your setup and con-
nections.
03
Loss of AC power; PF enabled and completed. The EC300 detected loss
of AC power, however the run was completed as programmed.
04
Loss of AC power; PF disabled and not completed. The EC300 detected
loss of AC power, the run was not completed since PF was not enabled.
05
Loss of AC power; run not timed. The EC300 detected loss of AC
power.
06-99
The EC300 detected an internal error condition. Contact Technical
Services for instructions.