background image

 

 9.  Reconnect the align/calibrate tool and close the top cover. The instrument is now converted to Class 1 

configuration.

 10.  Follow normal Align/Calibration procedure and make sure all tests pass.

 11.  Refer to the Class 3B Conversion Guide in the electronic documentation set for future reference.

12

DXR3xi Class 3B Conversion Guide

Converting DXR3xi from Class 3B to Class 1 configuration

Содержание DXR3xi

Страница 1: ...exposure to hazardous laser radiation l Set up a laser control area in the location where a Class 3B system is being used l When using or working in the vicinity of any Class 3B laser product each per...

Страница 2: ...eated Note For the complete laser safety information refer to the DXR Raman Instruments Site and Safety Guide A PDF copy is included in the electronic documentation set and in the online software Help...

Страница 3: ...ocks on the system l The connectors and the laser safety circuit controlling access to the Laser Control Area must be designed and built by personnel who have been trained in laser safety and in the o...

Страница 4: ...software 2 Remove the BNC interlock jumpers from the Interlock connectors on the instrument panel Properly store the jumpers in a labeled bag They will be used when you want to convert the instrument...

Страница 5: ...the inside of front ledge of the sample compartment turn the cover latch 90 degrees until it is in a horizontal position 6 Disconnect the align calibrate tool and set it aside 5 DXR3xi Class 3B Conver...

Страница 6: ...away from instrument 8 Take the laser interlock defeating dongle 085 0402xx CBL CLASS 3B OPERATION from the Class 3B Conversion Kit and install it to the interlock connector at the left side of sample...

Страница 7: ...is Class 3B system must wear laser safety eyewear that exceeds the required optical safety density for the laser wavelength in use Two laser safety goggles are provided in the DXR Class 3B Conversion...

Страница 8: ...on section in the DXRxi Raman Instrument User Guide for installation and align calibration instructions WARNING Avoid eye injury Never put anything except samples against the tip of the fiber optic pr...

Страница 9: ...he ports on the fiber optic launcher Do not twist the probe into a tight radius which can damage the fibers b Place the cap onto the end of the probe and the caps onto the ports on the fiber optic lau...

Страница 10: ...he external interlock BNC cables 5 Disconnect the align calibrate tool and set it aside 6 Make sure the two cover latches on the inside of front ledge of the sample compartment are in a horizontal pos...

Страница 11: ...seplate and slide it all the way to the back wall of the instrument Check that the cover sits properly without any gap around the sample area 8 Using a flat blade screwdriver to lock the two cover lat...

Страница 12: ...to Class 1 configuration 10 Follow normal Align Calibration procedure and make sure all tests pass 11 Refer to the Class 3B Conversion Guide in the electronic documentation set for future reference 1...

Отзывы: