Converting DXR3xi from Class 3B to Class 1
configuration
❖
To convert the DXR3xi from the Class 3B to the Class 1 configuration
1. Make sure the stage is positioned all the way to the back. Turn OFF the laser and the instrument. Close
OMNICxi software.
2. Carefully remove the Class 3B accessory from the instrument.
For example, to uninstall a fiber optic accessory:
a. Uninstall the fiber probe from the ports on the fiber optic launcher. Do not twist the probe into a tight
radius which can damage the fibers.
b. Place the cap onto the end of the probe and the caps onto the ports on the fiber optic launcher.
c. Remove the fiber optic launcher and replace with the proper wavelength edge filter into the
instrument.
3. Remove the interlock defeating dongle at the left side of sample area.
Properly store it in a labeled bag. (It will be used when you want to convert the instrument to Class 3B
configuration.)
9
DXR3xi Class 3B Conversion Guide
Converting DXR3xi from Class 3B to Class 1 configuration