9. Reconnect the align/calibrate tool onto the instrument.
10. Properly store the whole sample area cover in the lab.
11. The instrument is now converted to Class 3B configuration. Follow normal Align/Calibration procedure
and make sure all tests pass.
WARNING
Avoid eye injury.
Before turning on the laser, each person using or working in the vicinity of this
Class 3B system must wear laser safety eyewear that exceeds the required optical
safety density for the laser wavelength in use.
Two laser safety goggles are provided in the DXR Class 3B Conversion kit for 190-532 nm laser and 633-
1064 nm laser, respectively. If you need to order additional goggles please contact our local field
services or sales, or visit our web site at
http://www.thermo.com/spectroscopy
the appropriate goggles.
7
DXR3xi Class 3B Conversion Guide
Converting DXR3xi from Class 1 to Class 3B configuration