1
Basic Operations
Removing a Column from the GC/MS System
Thermo Scientific
DSQ II Hardware Manual
15
Removing a Column from the GC/MS System
To remove a column from the GC/MS system, you will need:
• Gloves, clean, lint- and powder-free
• Wrench, open-ended, 5/16-in.
• Wrench, open-ended, 7/16-in.
• Wrench, open-ended, 6 mm
1. Shut down the DSQ II.
a.
Shut down and vent the DSQ II (see
“Shutting Down the GC/MS System”
). Lower the oven, injector, and transfer line temperatures to 30 °C and allow
them to cool before continuing.
b. Once the DSQ II is cooled down, power off the TRACE GC Ultra.
2. Remove the column from the transfer line.
a.
Unscrew the transfer line nut.
b. Remove the column from the transfer line.
Figure 6.
Transfer Line Components
WARNING - BURN HAZARD:
The oven, injector, and transfer line are hot. Allow them to
cool to room temperature before touching them.
1
Septum
3
Transfer Line Union
2
Transfer Line Nut
4
Transfer Line
2
1
3
4
Содержание DSQ II
Страница 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Страница 4: ......
Страница 8: ......
Страница 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Страница 182: ...4 Vacuum System and Gas Inlets Gas Inlet Components 166 DSQ II Hardware Manual Thermo Scientific ...
Страница 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Страница 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Страница 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Страница 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...