Compatible Sensors for Smart-Vue Pro Duo / Quatro Data Logger

 Appendix 1 - Troubleshooting  |   27

2. When the battery is removed / discharged completely. 

This is observed as sensor unreachable in Smart-Vue Pro 
data logger.

3. When the Emerald goes out of range from the 

Smart-Vue Pro data logger.This is observed as Sensor 
unreachable in Smart-Vue Pro data logger.

4. When the External Emerald PT100 is removed /

damaged.

5. When any other wired sensor is connected other than the 

external Emerald PT100.

Will the instant reading be displayed on the Smart-Vue 
Pro data logger when tapped on the screen after 
unpairing the wireless remote sensors from the Smart-
Vue Pro data logger?

Instant reading screen is not displayed on the Smart-Vue Pro 
data logger when the wireless remote sensors are unpaired 
from the data logger during data logging.

In Smart-Vue Pro web application, why is the remove 
option disabled for the wireless remote sensors in the 
sensors page and enabled in the data logger page of 
the Smart-Vue Pro web application?

The Emerald has a specific handling in the sense that it has a 
battery (as opposed to the other sensors) and moreover, it 
can be a data logger too and not only a passive sensor. The 
very same module can have the 2 different ways of working, 
data logger VS sensor (but not simultaneously). So that's why 
it is shown in the data logger section too. Emerald data logger 
feature support is for future use.

Can the remote sensors be unpaired from the 
Smart-Vue Pro data logger during data logging when 
the remote sensors are unreachable from the 
Smart-Vue Pro data logger?

Yes, the remote sensors which are unreachable from the 
Smart-Vue Pro data logger can still be unpaired from the 
Smart-Vue Pro data logger with an error message display on 
the Smart-Vue Pro data logger "Operation incomplete: 
Remote sensor only unpaired locally.Reset your remote 
sensor to finish unpairing." To reset the remote sensor, 
disconnect the battery and connect the battery back again. 
then that particular remote sensor shall be visible in pairing list 
again.

During data logging, remote sensor is in out of range 
condition from the Smart-Vue Pro data logger and then 
the Emerald is unpaired from the Smart-Vue Pro data 
logger. Then how do we pair the remote sensor again 
when brought back within range of the Smart-Vue Pro 
data logger?

If the user tries to unpair the remote sensor which is in out of 
range from the Smart-Vue Pro data logger, a message is 
displayed on the Smart-Vue Pro data logger to reset the 
remote sensor to finish unpairing. So when the remote sensor 
is brought within range of the Smart-Vue Pro data logger, the 

user needs to reset the sensor (remove the battery and insert 
back again) to complete the unpairing process. Then user can 
pair the sensor again to the Smart-Vue Pro Duo/Quatro.

When the Emeralds are used as remote sensors, will 
the remote sensors store the data in the module during 
a Bluetooth network outage condition between Smart-
Vue Pro data logger and the Emeralds?

In this case, this is exactly like a disconnected Smart Sensor: 
the remote sensor is unreachable and this is a sensor failure. 
There is no local reading storage in the Emerald in this 
scenario: readings are lost during this out of coverage period 
(exactly like a disconnected Smart Sensor).

How many maximum number of Emeralds can we pair 
to Smart-Vue Pro Quatro / Duo (not starting data 
logging, only pairing)?

Maximum 4 can be paired to both the Smart-Vue Pro Quatro 
and Smart-Vue Pro Duo. 4 can be started with the Smart-Vue 
Pro Quatro, 2 with the Smart-Vue Pro Duo.

When the battery of Emerald is low, will there be a low 
battery alarm generated or a communication failure 
alarm after the battery is completely discharged?

A low battery alarm is generated in the Smart-Vue Pro web 
application for the Emerald.

When the Emerald is operating in the low battery 
condition, will it accurately read the temperature until 
the battery drops to zero completely or will there be a 
difference in the temperature read?

The low battery condition of Emerald does not effect the 
reading accuracy of the sensor.

In the Smart-Vue Pro web application, how is the 
remote sensor icon shown on the watch mode tile?

In Smart-Vue Pro web application watch mode, remote 
sensor shall be shown as temperature icon without any 
Bluetooth symbol. 

What happen for triple sensor when temperature / 
humidity sensor is removed? 

When the temperature / humidity probe is removed from the 
triple sensor, a sensor failure is detected for the temperature 
and humidity parameters while the CO

2

 parameter is in 

normal state logging the data as usual. 

How does the user start data logging for triple and 
differential pressure sensors in the Smart-Vue Pro web 
application?

Triple sensor and Differential pressure sensor data logging 
procedure is similar to any other wired sensor used in the 
Smart-Vue Pro web application.

Содержание Differential Pressure Sensor

Страница 1: ...e Sensors for Smart Vue Pro Duo Quatro Data Logger Emerald Wireless Remote Sensor Triple CO2 Temperature and Humidity Sensor Differential Pressure Sensor Supplemental User Guide 331942H01 Revision B 1...

Страница 2: ...med by qualified service personnel Material in this manual is for informational purposes only The contents and the product it describes are subject to change without notice Thermo Fisher Scientific ma...

Страница 3: ...s Remote Sensors 10 Wireless Remote Sensor Calibration 10 Key Specifications 11 Inserting the Battery into your Emerald Sensor 11 Activating an Emerald Wireless Remote Sensor 11 Mounting your Emerald...

Страница 4: ...Smart Sensor 22 Overview 22 Usage 22 Tube Placement 23 Key Specifications 24 Placement and Assembly 24 Velcro location on Extension box 25 Appendix 1 Troubleshooting 26 Appendix 2 Emerald Battery Life...

Страница 5: ...ct s enclosure that may be of sufficient magnitude to constitute a risk of electric shock to a person CAUTION This product contains two non rechargeable 3 6V lithium batteries Plugging the device into...

Страница 6: ...TION Do not use this device in explosive or hazardous environments CAUTION Do not expose your device to any extreme environment where the temperature or humidity is high Such exposure could result in...

Страница 7: ...and contact OCEASOFT technical services Battery Warning CAUTION This product contains a lithium battery Make sure you respect polarity when inserting batteries into OCEASOFT devices Reversing polarit...

Страница 8: ...subject to the following two conditions 1 This device may not cause interference 4 Certifications and Compliance 2 This device must accept any interference including interference that may cause undes...

Страница 9: ...is in compliance with the EU Directive 2011 65 EU Restriction of the Use of Certain Hazardous Substances in Electronic and Electrical Equipment and the amendment of EU 2015 863 Do not dispose of this...

Страница 10: ...imum or lesser gain approved for the transmitter by Industry Canada To reduce potential radio interference to other users the antenna type and its gain should be so chosen that the equivalent isotropi...

Страница 11: ...d other relevant provisions of the Waste Electrical and Electronic Equipment Directive 2012 19 EU WEEE Directive Environmental Protection RoHS Compliance Respect local regulations concerning disposal...

Страница 12: ...potentially expose you to chemicals including lead which is known to the State of California to cause cancer For more information go to www P65Warnings ca gov Respect local regulations concerning disp...

Страница 13: ...remote sensor Triple CO2 Temperature and Humidity sensor Differential pressure sensor The focus in this document is mainly on physical hardware aspects of the Smart Sensors listed above Where to go f...

Страница 14: ...s with up to two active wireless remote sensors on Smart Vue Pro Duo or up to four on Smart Vue Pro Quatro For the Smart Vue Pro data logger to detect your nearby Emerald wireless remote sensor you mu...

Страница 15: ...remote sensor you must first activate the sensor as described here 1 Place the strong side of the magnet the side without the adhesive tape in the circular recess on the back of the Emerald sensor an...

Страница 16: ...Adhesive Tape You may also choose to mount the Emerald sensor using the provided adhesive tape Note Clean the magnet before using the adhesive tape on it Using Wireless Remote Sensors with your Data L...

Страница 17: ...o another Smart Vue Pro data logger cannot be paired 6 Tap the sensor and confirm the connection by tapping OK 7 Wait until pairing is complete which may take a few seconds Once pairing is successful...

Страница 18: ...place damaged cables Wired sensors may be replaced at any time whether data logging is running or not You must ensure that the existing and replacement sensors are identical in type The process is com...

Страница 19: ...y the Smart Vue Pro web application Wireless Remote Sensors in Smart Vue Pro Web Application This section provides additional information regarding the use of wireless remote sensors in the Smart Vue...

Страница 20: ...data logger via the data logger menu displays the wireless remote sensor with the Bluetooth logo as shown here Figure 15 Wireless remote sensor shown on data logger Wireless Remote Sensor in the Senso...

Страница 21: ...o and is displayed purely for future use Wireless Remote Sensor Battery Indication The wireless remote sensor battery percentage is shown under Smart Tracker Pro as below Figure 20 Wireless remote sen...

Страница 22: ...inside the incubator Remove the CO2 temperature and humidity sensor from inside the incubator during decontamination procedures which may expose the sensor to damaging heat environment CAUTION Failure...

Страница 23: ...90 RH non condensing for indoor use Contact your Thermo Fisher Scientific representative for details regarding sensor accuracy drift and calibration Placing the Sensor and Smart Vue Pro Data Logger Fo...

Страница 24: ...cable into the Smart Sensor extension and the Smart Vue Pro data logger 2 this cable provides power to the Smart Vue Pro data logger Figure 27 Power connections for triple sensor unit 5 Connect the S...

Страница 25: ...uting sensor cables avoid direct contact with or close proximity of sensor cabling with any high voltage wiring Cabling should be placed with no less than a minimum of 2 inches 50 mm distance from hig...

Страница 26: ...placed outside the cleanroom so that people know it is safe to enter as shown in these two examples 1 High pressure cleanrooms such as vaccine and organ transplant laboratories in which the ambient pr...

Страница 27: ...he display A common way to use this sensor is therefore to plug the tube onto the high connector and under normal conditions obtain positive values when monitoring high pressure cleanrooms and negativ...

Страница 28: ...te or connector on the data logger 2 Clip the Smart Sensor extension firmly onto the bottom of the Smart Vue Pro data logger Figure 38 Slide the smart sensor extension onto Smart Vue Pro data logger 3...

Страница 29: ...Compatible Sensors for Smart Vue Pro Duo Quatro Data Logger Differential Pressure Smart Sensor 25 Velcro location on Extension box Figure 39 Velcro location...

Страница 30: ...in the Smart Vue Pro web application What should I do During pairing the Smart Vue Pro data logger downloads calibration parameters from the wireless remote sensor If wireless remote communication fa...

Страница 31: ...e user tries to unpair the remote sensor which is in out of range from the Smart Vue Pro data logger a message is displayed on the Smart Vue Pro data logger to reset the remote sensor to finish unpair...

Страница 32: ...a Logger Does the Emerald device send a low battery alert to the data logger Emerald low battery indication is not shown on Smart Vue Pro Duo Quatro data logger but Emerald low batery status level are...

Страница 33: ...conditions Wireless remote communications Bluetooth connections consume battery power However it is worth noting that the Smart Vue Pro data logger reads Bluetooth advertising frames from the Emerald...

Страница 34: ...thermofisher com WEEERoHS Conformit WEEE Questo prodotto deve rispondere alla direttiva dell Unione Europea 2012 19 EU in merito ai Rifiuti degli Apparecchi Elettrici ed Elettronici WEEE marcato col...

Страница 35: ...e 33 2 2803 2180 Germany international 49 6184 90 6000 Germany national toll free 0800 1 536 376 Italy 32 02 95059 552 Netherlands 31 76 579 55 55 Nordic Baltic CIS countries 358 9 329 10200 Russia 7...

Отзывы: