5 - Troubleshooting
Troubleshooting Tables
Problem Possible
Reason Advice
Sections fold or
crumple
1. Specimen is too warm.
Lower specimen temperature
2. Knife holder too warm.
Lower chamber temperature.
Tip – use freezer spray on clamp plate and blade
to confirm.
3. Anti-Roll plate too low.
Raise Anti-Roll towards specimen.
4. Anti-Roll plate and/or
clamp plate dirty.
Clean with absolute alcohol & dry
thoroughly.
Section rolls up under
the Anti-Roll plate
1.Specimen is too cold.
Raise specimen temperature.
Specimen curls after
lifting ARP
1. Anti-Roll plate & clamp
plate too warm.
Lower chamber temperature
Tip – use freezer spray on clamp plate and blade
to confirm.
2. Static electricity in
chamber.
3. Blade blunt.
Change blade.
Sections tear or crack
1. Specimen too cold.
Raise temperature.
2. Blade damaged or dirty.
Change blade.
3. Specimen frozen too
rapidly or specimen overly
large.
Specimen and section
chatter
1. Knife holder not correctly
clamped.
Check & tighten all stages of knife holder.
2. Blade incorrectly clamped.
Blade clamping force can be increased by
tightening clamp screw at the rear of the top
stage.