88
Sectioning Procedures
Cutting Movement and Retraction
Front
Rear
Lateral view into instrument
Notes
The retraction prevents the surface of the specimen being damaged during the return stroke of the specimen head.
The retraction can be switched off - see "Switching the Retraction On - Off".
The retraction distance required is 20 microns and cannot be changed.
Trimming and First Cuts
After the specimen orientation and the blade angle have been set, trimming can be carried out using the trimming
function.