B–8
Thermo Scientific Cryofuge 8 / 16 Centrifuges and Heavy Duty Centrifuges
Rotors
HAEMAFlex 12 Rotor
with Double Blood Bag Buckets
Technical Data
Cryofuge 16 / Cryofuge 16 Heavy Duty
Centrifuge Voltage
200, 208, 220,
230, 240 V
380, 400, 415 V
Rotor Capacity
12 x 1000 ml
12 x 1000 ml
Maximum Speed n
max
4
700 rpm
4
700 rpm
Maximum RCF-Value at n
max
7
187 x g
7
187 x g
K-Factor at n
max
7
909
7
909
Radius max. / min.
291 mm / 146 mm
291 mm / 146 mm
Angle
90°
90°
Maximum Speed at 4 °C
4
600 rpm
4
600 rpm
Sample Temperature
at Max. Speed
(Ambient temperature of 23 °C,
run time 90 minutes)
50 Hz: 5 °C
60 Hz: 2 °C
50 Hz: 5 °C
60 Hz: 2 °C
Aerosol-tight
No
No
Rotor Body
Max. Cycle Number
50 000
Buckets
Max. Cycle Number
1
at
7187 x g / 4700 rpm
5000 x g / 3921 rpm
4000 x g / 3507 rpm
2500 x g / 2772 rpm
14 000
31 000
50 000
137 000
Weight
25.6 kg
Rotor Body (empty)
Windshield Lid
Windshield Bowl
18.5 kg
1.7 kg
5.4 kg
1
Cycle numbers stated are only valid, if the according speed (or rcf) is always set for centrifugation. If different speed (or rcf) settings are used, the
smallest cycle number stated is valid.
Table B–8:
HAEMAFlex 12 Rotor with Double Blood Bag Buckets