III–3
Thermo Scientific Cryofuge 8 / 16 Centrifuges and Heavy Duty Centrifuges
Operation
3. 2. Rotor Installation
CAUTION
An an improper installation of the rotor can cause severe damage to the rotor and the centrifuge.
Make sure the rotor is locked properly into place before operating the centrifuge.
Use the rotor locking tool (20360104) clockwise to check if the rotor locking screw is tightened to the firm stop.
CAUTION
If you have not installed the rotor correctly, do not operate it.
Make sure that the rotor is not damaged. Do not use damaged rotors.
Make sure that the centrifugation chamber is free from objects.
1. Open the centrifuge door.
If necessary remove dust and foreign objects from the chamber.
Wipe the drive shaft with a clean cloth. Clean the threads of the drive shaft to prevent possible damage to rotor or centrifuge.
Apply one drop of gresae (75003786) to the drive shaft. Wipe the rotor hub from the bottom side of the rotor with a clean
cloth.
2. Hold the windshield bowl at the windshield knob with one hand and at the rim with the other hand.
Put the windshield bowl carefully on the drive shaft.
For some windshield bowls it may be easier to handle them with 2 people.
3. Align the square symbol on the windshield knob to the square end on the drive shaft.
➀
Drive shaft
➁
Windshield knob
➂
Square symbol on windshield
knob
Figure III–2:
Align windshield knob and drive shaft
4. Identify the circular marking on the rotor body.
Figure III–3:
Identify the circular marking, e.g. HAEMAFlex 8 and 12
The hole for the locating pin of the windshield bowl is on the bottom side of the rotor body below the circular marking.
Hold the rotor body over the drive shaft and align the circular marking to the locating pin on the windshield bowl. Put the
rotor body carefully on the drive shaft. Turn the rotor body against the windshield bowl until the locating pin engages into
the rotor body.