Troubleshooting
Table 3. Furnace Troubleshooting.
Troubleshooting procedures involve working with high voltages
which can cause injury or death. Troubleshooting should only be
performed by trained personnel. This section is a guide to
troubleshooting controller and furnace problem.
Problem
Solution
Absence of current and presence of voltage on
element input terminal.
Heating element is cracked or open.
Replace broken element.
Heating element curls towards wall
1. Shut off immediately.
2. After the furnace cools, relocate the
element farther from wall.
Controller actual furnace temperature display
decreases when furnace is heating
Check thermocouple polarity. Red is always
negative (Refer to Section Troubleshoot-
ing).
Содержание BF51634BC-1
Страница 1: ......
Страница 6: ...List of Figures ii Thermo Scientific...
Страница 20: ...Safety Notes Standards and Directives 1 14 Thermo Scientific...
Страница 48: ...Communication Option Auto Tuning the Controller 7 10 Thermo Scientific...
Страница 52: ...Excess Temperature Option B Models Excess Temperature Controller Operation 8 4 Thermo Scientific...
Страница 64: ...Troubleshooting 10 4 Thermo Scientific...
Страница 65: ......
Страница 66: ......
Страница 69: ...Thermo Scientific 12 1 12 Error Codes This chapter is not applicable...
Страница 70: ...Error Codes 12 2 Thermo Scientific...
Страница 72: ...WEEE Compliance 13 2 Thermo Scientific...
Страница 73: ...Thermo Scientific 14 1 14 14 Spare Parts and Accessories This chapter is not applicable...
Страница 74: ...Spare Parts and Accessories 14 2 Thermo Scientific...
Страница 76: ...Device Log 15 2 Thermo Scientific...