Pretreatment system
Technical Specifications | 5
Technical Specifications
Note:
Check at regular intervals the quality of your feed-water.
Feed-water requirements
Source
Potable tap water
Silt density index (SDI)
<5
Turbidity
<1 NTU
Conductivity (reference
tempe25°C)
1500 μS/cm (670 Ω·cm)
Chlorine
Max. 3 ppm
Iron content
Max. 0.5 ppm
pH range
4 - 11
Temperature
+2°C - +35°C
Pressure
2 - 6.9 bar / 29 - 100 PSI (at a pressure >6.9 bar / 100 PSI a pressure reducer must
be installed upstream of the system.
Feed-water supply
¾”
NPT (adapts to ¼” NPT male or female)
Tubing Connections
Inlet Tubing
Tube ¼” outer diameter
Outlet Tubing
Tube ¼” outer diameter
Accessibility to Pretreatment system
Space to the Left & Right of Unit
15 cm / 6 inch
Space to the Front of the Unit
30 cm / 12 inch