Perform ROI/uniformity, background, dye, and
normalization calibrations for the TaqMan
™
Array Card
block
To calibrate the plate blocks, see “Perform ROI/uniformity, background, and dye calibrations for plate
blocks”
This section covers procedures specific to the TaqMan
™
Array Card block. See the following sections
for additional information:
• Transfer calibration results (
• Transfer data files from calibrations (
• Troubleshoot calibration failure (
To troubleshoot normalization calibration, see “Troubleshoot normalization calibration failure”
.
Calibration workflow
Perform an ROI/uniformity calibration
You are automatically prompted to perform background calibration.
▼
Perform a background calibration
Perform any time that ROI/uniformity calibrations are current.
▼
Perform system dye calibrations
Perform any time that ROI/uniformity and background calibrations are current.
▼
Perform normalization calibration
Perform after a dye calibration.
▼
(
Optional
) Perform custom dye calibrations
)
Perform any time that ROI/uniformity and background calibrations are current.