Review calibration status
The calibration status can be reviewed in the QuantStudio
™
Design and Analysis Software v2. The
instrument must be added to the
Manage Instruments
list.
1.
In the home screen, click
Manage Instruments
.
To access instruments from a different screen, click
System
4
Instruments
.
2.
In the
Summary
tab for the instrument, the calibration status is displayed in the
Calibration
Status
column.
Export a calibration report
The calibration report is exported as a TXT file.
1.
In the home screen, tap
(Settings)
4
Maintenance and service
4
Calibration
4
Status and
reminders
.
2.
Tap
Export
, then select a destination.
•
Connect
•
Network Drive
•
USB Drive
The connectivity status of each location is displayed.
3.
Tap
Done
.
Perform ROI/uniformity, background, and dye calibrations
for plate blocks
To calibrate the TaqMan
™
Array Card block, see “Perform ROI/uniformity, background, dye, and
normalization calibrations for the TaqMan