1
Introduction
Specifications
8
Accela RI Detector Hardware Manual
Thermo Scientific
Fuse size
• For 110/115 V ac: T3.15 A, 5×20 mm
• For 230 V ac: T1.6A, 5×20 mm
Operating humidity
5% to 95% non-condensing relative humidity
Method of detection
Deflection
Refractive index range
1.00 to 1.75 RIU
Attenuator range
0.25 to 512 μRIU
Linearity range
600 μRIU
Noise
= 2.5 × 10
–9
RIU (Response:1.5 seconds)
Rise time
0.1, 0.25, 0.5, 1, 2. 3. 6 seconds
Polarity
Positive/Negative
Zero adjustment
Optical zeroing, automatic zeroing
Automatic zeroing range
± 10 × 10
–5
RIU
Resolution of autozeroing
± 10 × 10
–9
RIU
Offset adjustable range
0 to 500 mV (same with Integrator output sensitivity)
Resolution of offset
10 mV (same with Integrator output sensitivity)
Integrator output
0 to 1V/FS (Sensitivity: 512 × 10
–6
RIU/FS)
Recorder output
0 to 10 mV/FS
External communication
Ethernet
Event marker input
Contact: = DC 5V, 0.1 A
• Purge on/off
• Automatic zeroing on/off
• Run
Signal output
Contact: = 5V, 0.1 A Ready
Signal input
Run, Autozero, Purge
Temperature control
30 to 50 °C (1 °C increment) or off
Flowcell volume
8 μL
Maximum flow rate
10 mL/min (mobile phase: deionized water)
Pressure rating
517 kPa (75 psi)
Liquid-contact materials
316 stainless steel, Teflon™, Quartz optical glass
Table 2.
Accela RI Detector specifications (Sheet 2 of 2)
Item
Specification