To defrost the storage tank, remove the product and LN
2
from the tank.
Allow the tank to warm at room temperature.
Caution
Do not attempt to speed up LN
2
evaporation, or frost removal, by
blowing direct heat. Avoid using a heater or blowing compressed air into
the storage tank.
▲
To defrost the vent port, open the lid and allow to warm at room
temperature. See Caution above.
The opening of the storage container can be loosely covered while the vent
port thaws.
After the port has thawed, remove all moisture from the vent port area
before returning the tank to service.
7007400 Cryoplus 4-3
Thermo Scientific
Section 4
Maintenance
Defrosting the
Storage Tank
Defrosting the Vent
Port