4-2
7007400 Cryoplus
Thermo Scientific
Section 4
Maintenance
3. Clean and assemble in reverse order of disassembly, paying careful
attention to exploded view provided (Figure 4-1) for identification
and placement of parts.
Note
Use thread sealant, not Teflon tape. Tape could get into valve.
▲
4. After maintenance, operate the valve a few times to ensure proper
opening and closing.
Protective clothing and gloves should be worn whenever cleaning the inside
of this unit. Follow established laboratory procedures. Allow the interior
of the unit to warm to ambient and use an appropriate germicide.
Freezing chamber
The interior of the CryoPlus unit is made of high quality stainless steel
and should not be cleaned with any cleanser containing chlorine.
Exterior cabinet
Wash the outside of the cabinet with a mild detergent.
Figure 4-1.
Exploded View of Valve
Housing
Spring Washer
Coil
Solenoid Base Sub-assembly
Core Assembly with Rider Ring
Core Spring
Body Gasket
Valve Body
General Cleaning
Valve Seat