| Cleaning and Disinfection
9-4
|
Model 3900 Series
Forma
Chapter 9
Cleaning Operation Panels
The operation panels are moisture-sensitive.
Do not spray or wipe with cleaner.
Clean using a dry cloth of 100% micro fibre.
Wipe / Spray Disinfection
The manual wipe/spray disinfection is carried out in three stages:
Predisinfection
Cleaning
Final disinfection
When using such disinfectants, avoid open flames or exposure to excessive heat during
the entire disinfection process!
Use such disinfectants only in adequately ventilated rooms.
After the disinfectant has been allowed to react, wipe the cleaned device components
thoroughly dry.
Observe safety regulations to avoid fire and explosion hazard caused by
alcohol-containing disinfectants.
Predisinfection
Remove all samples from the work space and store them in a safe place.
Spray disinfectant onto the surfaces of the work space and of the accessories or wipe the
surfaces clean using disinfectant.
Alcoholic disinfectants
Disinfectants having an alcohol content of more than 10% may form, in
combination with air, easily combustible and explosive gas mixtures.
Chloride-containing disinfectants
Do not use chloride-containing disinfectants.
Содержание 3911
Страница 10: ... Safety Notes 1 6 Model 3900 Series Forma Chapter 1 ...
Страница 20: ... Installation 4 6 Model 3900 Series Forma Chapter 4 ...
Страница 32: ... Operation 7 2 Model 3900 Series Forma Chapter 7 ...
Страница 44: ... Cleaning and Disinfection 9 6 Model 3900 Series Forma Chapter 9 ...
Страница 50: ... Disposal 11 2 Model 3900 Series Forma Chapter 11 ...
Страница 70: ... Service 13 14 Model 3900 Series Forma Chapter 13 Honeywell Truline Configuration Record ...
Страница 71: ...Forma Model 3900 Series 13 15 Service Chapter 13 ...
Страница 72: ... Service 13 16 Model 3900 Series Forma Chapter 13 ...
Страница 73: ...Forma Model 3900 Series 13 17 Service Chapter 13 ...
Страница 74: ... Service 13 18 Model 3900 Series Forma Chapter 13 ...