Forma
Model 3900 Series |
5-3
Product Description |
Chapter 5
The CO
2
source must be regulated at a pressure level of 15, ±5 psig.Higher pressure
levels may damage the CO
2
control system. The user should determine the most
economical pressure level, between 10 psig and 20 psig appropriate for the desired CO
2
percentage in the chamber. Use only sufficient pressure to maintain recovery time after
door openings.
Figure 5-3
Inlet Connections
To connect the CO
2
supply (Figure 5-3):
1. Connect the CO
2
tubing to the ¼” hose fitting installed in the CO
2
inlet.
2. Check the tubing connection for leaks.
H
2
O inlet
CO
2
inlet
Do not leave unit door open when using CO
2
.
Raised levels of CO
2
can lead to asphyxiation, only use in well ventilated rooms.
If CO
2
inlet pressure level is higher than 15 +-5psig, liquid CO
2
might enter the
unit and damage the CO
2
control system, valves, tubing and filters.
Содержание 3911
Страница 10: ... Safety Notes 1 6 Model 3900 Series Forma Chapter 1 ...
Страница 20: ... Installation 4 6 Model 3900 Series Forma Chapter 4 ...
Страница 32: ... Operation 7 2 Model 3900 Series Forma Chapter 7 ...
Страница 44: ... Cleaning and Disinfection 9 6 Model 3900 Series Forma Chapter 9 ...
Страница 50: ... Disposal 11 2 Model 3900 Series Forma Chapter 11 ...
Страница 70: ... Service 13 14 Model 3900 Series Forma Chapter 13 Honeywell Truline Configuration Record ...
Страница 71: ...Forma Model 3900 Series 13 15 Service Chapter 13 ...
Страница 72: ... Service 13 16 Model 3900 Series Forma Chapter 13 ...
Страница 73: ...Forma Model 3900 Series 13 17 Service Chapter 13 ...
Страница 74: ... Service 13 18 Model 3900 Series Forma Chapter 13 ...