Service
Remove Software Lockout
1. Press the Advance and Infinity keys at the same time and hold them for about six
seconds. The word “Fcty” (factory) will appear in the bottom display. If numbers in the
bottom display begin to scroll up or down, the keys have not been pressed
simultaneously. Try again.
Figure 13-1
Displays
2. Press the Up Arrow until “LoC” (lock) appears in the upper display. The word “Fcty”
will remain in the lower display (Figure 13-1).
3. Press the Advance key to scroll through the menus as follows:
Service must be performed by qualified service personnel only!
De-energize all potential sources of energy to this unit and lockout/tagout their
controls.
Lock
Factory
Press
(6 sec)
Lower display
Upper display
Keystrokes
LoC.o
1
Change to 3 = unlocked
LoC.P
1
No changes required
PAS.E
1
No changes required
rLoC
1
Change to 5 = unlocked
SLoC
1
Change to 5 = unlocked
Содержание 3911
Страница 10: ... Safety Notes 1 6 Model 3900 Series Forma Chapter 1 ...
Страница 20: ... Installation 4 6 Model 3900 Series Forma Chapter 4 ...
Страница 32: ... Operation 7 2 Model 3900 Series Forma Chapter 7 ...
Страница 44: ... Cleaning and Disinfection 9 6 Model 3900 Series Forma Chapter 9 ...
Страница 50: ... Disposal 11 2 Model 3900 Series Forma Chapter 11 ...
Страница 70: ... Service 13 14 Model 3900 Series Forma Chapter 13 Honeywell Truline Configuration Record ...
Страница 71: ...Forma Model 3900 Series 13 15 Service Chapter 13 ...
Страница 72: ... Service 13 16 Model 3900 Series Forma Chapter 13 ...
Страница 73: ...Forma Model 3900 Series 13 17 Service Chapter 13 ...
Страница 74: ... Service 13 18 Model 3900 Series Forma Chapter 13 ...