Section 7
Operation
Standard Model Only
1. Place the HEAT "On-Off"” switch in the "On" position.
2. Place the WATTAGE selection switch in the “1400” watts position for
fast chamber heat-up.
Allow the chamber temperature to rise to within 5°C of the desired
temperature; then select the appropriate heater wattage.
Desired Chamber
Wattage Setting
Temperature °C
350
37 to 45
700
46 to 60
1400
61 to 70
Low Temperature Model Only
1. Place the COOL-HEAT switch in the switch positions indicated below
to maintain the desired chamber temperature.
Desired Temp.
Switch Position
Range
COOL 5°C to ambient plus 10°C
HEAT
Ambient plus 10°C to 70°C
2. When the HEAT position is employed, place the WATTAGE
selection switch in the “1400” watts position for fast chamber heat-up.
Allow the chamber temperature to rise to within 5°C of the desired
temperature; then select the appropriate heater wattage.
Desired Chamber
Wattage Setting
Temperature °C
350
37 to 45
700
46 to 60
1400
61 to 70
30 cu ft Incubators
7-1
Thermo Scientific