Section 6
Installation
Location: To assure proper ventilation, allow a minimum of 4 inches of
clearance between the rear, top, and sides of the incubator and adjacent
walls.
Caution
An exhaust vent is located at the top right side of the cabinet.
it is not adjustable and should never be obstructed.
s
If two or more incubators are positioned side by side, allow a minimum
of 8 inches between cabinets. Adjust the front leveling feet of the
incubator so that the front is higher than the rear. This will assist in
door closing. Choose a site free from rapidly changing ambient
temperature conditions or one which will experience high rises in ambient
temperature during the summer months.
Radiators, air-conditioning outlets, other ventilating system outlets, and
drafts can affect the operation of the incubator by a sudden inrush of air
that is at a temperature different than operating conditions.
Electrical Connections:
Important, please read carefully.
Caution
Be sure that the power supply is of the same voltage as specified
on the nameplate.
s
The incubator is supplied with mains power leads located beneath the
control compartment cover, and must be permanently connected to an
electrical source by a qualified electrician.
Warning
The National Electrical Code should be observed for proper
fusing and the size of the service wires. For personal safety, this apparatus
must be properly grounded.
s
A power cord is provided at the rear of the cabinet for the Convenience
Outlet, 15 amps maximum, located inside the chamber.
30 cu ft Incubators
6-1
Thermo Scientific
Содержание 30M
Страница 28: ...8 6 30 cu ft Incubators Thermo Scientific ...
Страница 29: ...30 cu ft Incubators 8 7 Thermo Scientific ...
Страница 30: ...8 8 30 cu ft Incubators Thermo Scientific ...
Страница 31: ...30 cu ft Incubators 8 9 Thermo Scientific ...
Страница 33: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com ...