30 cu ft Incubators
6-3
Thermo Scientific
Section 6
Installation
WARNING: SAFETY PRECAUTIONS
1. DO NOT place any explosive, combustible, or flammable materials in
the chamber.
2. DO NOT place sealed containers in the chamber. Sealed
containers, filled with materials, do not provide room for expansion
and can develop dangerous vapor pressure as the temperature
increases.
3. Avoid spillage of large volumes of liquids.
4. DO NOT evaporate noxious fumes.
Low Temperature Model Only
A plastic container is provided to collect condensate from the cooling coil.
A piece of 7/16" I.D. plastic tubing is used to connect the condensate drip
tube (located at the rear center of the chamber ceiling) to this container.
Note
The liquid level, in the container, should be checked daily to
prevent overflow.