Chapter 3
SPECIFICATIONS
3-2
MR23i
Thermo Fisher Scientific
The maximum rotational energy is not systematically that obtained
with the maximum permissible load on the drive shaft.
Acceleration rate control
The selection of the rate acts on the acceleration time between 0
and 2 000 rpm. Above 2 000 rpm a max rate is applied. However,
the total acceleration time depends upon the inertia of the rotor.
Range 10 rates (0 to 9)
Duration (± 15%) 0 to 90% of maximum speed (slope 9)
1
:
Braking rate control
Maximum braking rate is applied until the rotor reaches 2 000 rpm
below which the chosen rate comes into effect. Above 2 000 rpm
there would be no gain in quality of separation if a slower rate of
braking could be applied.
Range 10 rates (0 to 9)
Duration (± 15%) from maximum speed to 0 rpm (slope 9)
Integrator
Range
100 to 281 x 10
12
m.
rd
2
/s
Step
1 m. rd
2
/s
Acceleration rates
10
Braking rates
10
Temperature
Range
-8°C to +40°C
Step 1°C
Accuracy
± 1.5°C
Typical
performance
-5°C at 4400 rpm, 3074 x g (4 x 200 ml
swing-out)
1°C at 23000 rpm, 34780 x g (12 x 1.5 ml angle)
Max. density
1,2 g/ml
Max. load on drive
shaft
5.4 kg
Max. energy
26900 J
DRM 6.14
45 sec.
AM 38.15
90 sec.
AM 2.23
35 sec.
AM 50C.13
260 sec.
AM 2.18
55 sec.
AM 50.14
120 sec.
AM 2.19
50 sec.
AM 100.13
150 sec.
FAJM 2.15C
125 sec.
SWM 180.5
30 sec.
AM 10.17
70 sec.
MTM 6.4
30 sec.
1
Typical acceleration and braking times with an empty rotor.
Содержание 11174729
Страница 1: ...P N 50116911 1 Analyze Detect Measure Control Instruction Manual MR 23i ...
Страница 2: ......
Страница 3: ...PN 50116911 1 Issued 06 08 Instruction Manual Jouan MR 23i ...
Страница 4: ......
Страница 14: ...Chapter 2 INSTALLATION PROCEDURE 2 6 MR23i Thermo Fisher Scientific ...
Страница 26: ...Chapter 4 OPERATING PRINCIPALES 4 6 MR23i Thermo Fisher Scientific ...
Страница 54: ...Chapter 6 HAZARDS PRECAUTIONS AND LIMITATIONS OF USE 6 18 MR23i Thermo Fisher Scientific ...
Страница 58: ...Annex A GUARANTEE TERMS A 2 MR23i Thermo Fisher Scientific ...
Страница 62: ...Annex B Declaration of conformity B 4 Heraeus Biofuge Primo R Thermo Fisher Scientific ...
Страница 64: ...Index I 2 MR23i Thermo Fisher Scientific ...
Страница 65: ......