Model 916/917_________________________________________________________________________________Service
5-2
No Temperature Display
•
Power line cord disconnected or not properly installed
•
External power circuit breaker tripped/open
•
Main power switch OFF
•
24 volt transformer fuses open.
Symptom
Possible Cause
Chamber Temperature Deviates from Set Point
•
Too much warm product added
•
Door open too long
•
Inadequate air circulation
•
Calibration
•
Dirty condenser
•
High ambient temperature
Too Much Frost Build-Up
•
Door not properly sealed.
Freezer not Being Refrigerated (unit is receiving power)
•
Compressor thermal overload open
•
Loss of refrigerant in either system
•
Defective compressor(s)
•
Defective temp control
•
Defective high pressure cut-off
Display Problems in General
•
Defective temp control board.
Condenser Hot Light On
•
Clogged air filter
•
Clogged condenser
•
Fan failure
5.2 Troubleshooting Chart
Содержание 916
Страница 20: ...7 1 Model 916 917_____________________________________________________________________________Parts List ...
Страница 21: ...Model 916 917_____________________________________________________________________________Parts List 7 2 ...
Страница 22: ...7 3 Model 916 917____________________________________________________________________________Parts List ...
Страница 23: ...Model 916 917_______________________________________________________________________________Parts List 7 4 ...
Страница 24: ...7 5 Model 916 917_____________________________________________________________________________Parts List ...
Страница 25: ...7 6 Model 916 917_____________________________________________________________________________Parts List ...
Страница 26: ...7 7 Model 916 917_____________________________________________________________________________Parts List ...
Страница 27: ...8 1 Model 916 917___________________________________________________________________Refrigeration Schematics ...
Страница 28: ...8 2 Model 916 917___________________________________________________________________Refrigeration Schematics ...
Страница 29: ...8 3 Model 916 917___________________________________________________________________Refrigeration Schematics ...
Страница 30: ...Model 916 917___________________________________________________________________Electrical Schematics 9 1 ...
Страница 31: ...Model 916 917___________________________________________________________________Electrical Schematics 9 2 ...
Страница 32: ...Model 916 917___________________________________________________________________Electrical Schematics 9 3 ...
Страница 33: ...Model 916 917__________________________________________________________________Electrical Schematics 9 4 ...
Страница 34: ...Model 916 917__________________________________________________________________Electrical Schematics 9 5 ...
Страница 35: ...Model 916 917__________________________________________________________________Electrical Schematics 9 6 ...
Страница 39: ...Millcreek Road P O Box 649 Marietta Ohio 45750 U S A Telephone 740 373 4763 Telefax 740 373 4189 ...