background image

CODE REMATCH 

1.

 

Wireless Transmitter has been synchronized with the Thermometer/Receiver at ex-factory setting and serialized to match.  If rematch is 

needed, turn off the Thermometer/Receiver, press and hold [MODE] and turn on the thermometer after 5 seconds. Rematch mode will 

start and “Syn” icon will appear in the display. 

2.

 

Turn off the Wireless Transmitter press and hold [ ]  for 3 seconds; Blue LED light above power button will flash 2 times.  Now turn on 
the Wireless Transmitter by press and hold [ ] for 5 seconds to start rematch. 

3.

 

The units will restore the new code automatically and return to NORMAL <Current/Min/Max> display mode. 

NOTE 

1.

 

If the Thermometer/Receiver is not in use for a long period of time remove the batteries from battery compartment to avoid battery 

leakage. 

2.

 

Do not operate the Thermometer/Receiver in environmental temperature lower than 0

o

C/32

o

F or higher than 50

o

C/122

o

F otherwise 

incorrect readings or damage to the Thermometer/Receiver may result.  The Thermometer/Receiver should never be placed in an 

appliance for temperature monitoring, only the Probe or Wireless Transmitter. 

 

 

Measuring Range 

0~50°C / 32~122°F (thermometer internal sensor) 

-50°C~70°C  / -58~158°F (External probe sensor) 

-40°C~50°C  / -40~122°F (External wireless sensor) 

Display Accuracy 

±0.5°C / ±0.9°F in -20 ~50°C / -4~122°F, ±1.0°C / ±1.8°F rest range 

Display Resolution 

0.1° 

Display Reading Update 

10 seconds thermometer internal sensor; 30 seconds external wireless 

sensor 

Wireless Sensor Distance 

50m (open field), 10m (inside appliance) 

Battery 

1.5 volt, type AAA x 4 pieces 

Display Size 

(W) 79mm x (H) 66mm 

Thermometer Receiver Battery Life  ~ 15,000 hours in probe mode 

~ 4,000 hours in wireless mode 

Wireless Transmitter Battery Life 

~ 6,000 hours 

Accessories 

1 each, external probe sensor, 1m cord 

4 each, AAA batteries 

1 each, external wireless sensor 

 

ERROR SYMBOLS 

SYMBOL 

DESCRIPTION 

ACTION 

 

Transmitter low battery voltage 

Replace battery 

 

Receiver low battery voltage 

Replace battery 

LL.L 

1.

 

Sensor open circuit 

2.

 

The reading exceeds lower range of 

thermometer (-50°C/-58°F) 

1.

 

Return the thermometer for evaluation/repair 

2.

 

Adjust for reading to be in range of lower limit 

HH.H 

1.

 

Sensor open circuit 

2.

 

The reading exceeds upper range of 

thermometer (70°C/158°F) 

1.

 

Return the thermometer for evaluation/repair 

2.

 

Adjust for reading to be in range of upper limit 

Con 

Receiver is searching the signal from the 

transmitter 

1.

 

Turn on the transmitter 

2.

 

No action required (searching when powered on) 

Syn 

Code rematch 

No action required 

FAI 

Connection failure 

1.

 

Review product is in valid distance range 

2.

 

Power device on/off 

 

LIMITED WARRANTY 

1 year from date of distributor purchase due to manufacture defect of product functionality. 

 
SDS – 

https://www.thermcoproducts.com/sds.html  

 

 

 