FL-B101D40PF Power factor (D0109) Manual

Page 3

The FL-B101D40PF is an intelligent bargraph meter with a hier-

archical software structure designed for easy programming and 

operation, as shown below in the software logic tree.

Software Logic Tree

Peak

[PEAK]

Reset

PEAK

Reset

VALY

Valley

[VALY]

[orAn]

[rEd]

NOTE: Will only 

appear if a tricolor 

display is installed

[orAn]

[rEd]

[orAn]

[rEd]

[orAn]

[rEd]

[orAn]

[rEd]

BARGRAPH COLOR 

PROGRAMMING MODE

See Page 9

Tricolor Bargraph

The tri-color bargraph is 

designed like a traffic light, to 

display either red, orange or 

green, but only one color at a 

time. The color to be displayed 

is selected in two ways.

The first step is to select the 

color to be displayed when the 

bar is “below” whichever set 

point is set to the lowest 

position.

The second step is to select the 

color to be displayed when the 

bar is above each specific 

setpoint, regardless of the order 

or position to which the set 

points are set.

However, if two or more 

setpoints with differently 

specified colors are positioned 

at the same set point value, the 

color specified for the set point 

with the highest identifying 

number will be displayed. When 

setpoints are set to the same 

value, the SP4 color overrides 

the SP3 color, the SP3 color 

overrides the SP2 color, and the 

SP2 color overrides the SP1 

color.

Set Setpoint 1

[SP1]

Delay-on-Make

[doM]

Delay-on-Break

[dob]

Relays 

Activation 

[rLYS] 
(h) High
 the 

relay 

energizes 

when the 

setpoint is 

exceeded.
(L) Low 

the relay

energizes 

below the 

setpoint. 

Setpoint 2

[SP2]

Setpoint 2

[SP2]

Setpoint 1

[SP1]

Setpoint 3

[SP3]

NOTE: [dom] [dob] 

Functions are only 

available for SP1 

0 to 9999

seconds

0 to 9999

seconds

Pump

[PUM]

Hysteresis

[hYSt]

Hysteresis

[hYSt]

Bar Color

[coL]

Hysteresis

[hYSt]

Hysteresis

[hYSt]

Setpoint 3

[SP3]

Setpoint 4

[SP4]

Setpoint 4

[SP4]

Operational Display

MAIN  MENU

PEAK & VALLEY 

VIEW & RESET

BRIGHTNESS AND

SETPOINT VIEW MODE

Bargraph Center Point

Display Selection [cto]

Select [on]

This branch will

only appear if

the analog

output option

is installed

Back to

Operational

Display

Digital Display

ON/OFF 

Selection [diSP]

Sub-menu MODE

Calibration Mode

Calibration

Input

Calibration

Output

Calibrate Analog 

Output Low

[cLo] 

Calibrate Analog

Output High

[chi]

Select the Digital

Reading at which

the Analog Output

Hi [chi] will occur

Select the Digital

Reading at which

the Analog Output

Lo [cLo] will occur

This branch will only appear 

if the analog output option 

is installed 

DIGITAL SPAN SELECTION

FOR ANALOG RANGE OUTPUT

See Page 6

DIGITAL DISPLAY ON/OFF

See Page 5

BARGRAPH CENTER POINT

DISPLAY MODE SELECTION

See Page 6

TWO POINT ANALOG OUTPUT

RANGE SETTING AND CALIBRATION

See Page 7

Set the Decimal

Point [dp]

Select 

[XX•XX]

 

Bargraph Display

Scale Lo Range

Setting [bLo]

Adjust to: 50

Bargraph Display

Scale hi Range

Setting [bhi]

Adjust to: 150

Display

Brightness [br]

1  Dimmest

2  Dim        

3  Bright     

4  Brightest

[X•XXX]

[XX•XX]

[XXX•X]

[XXXX•]

[XXXX]

DECIMAL POINT AND

BRIGHTNESS SELECTION

See Page 5

DIGITAL SPAN SELECTION

FOR BARGRAPH DISPLAY

See Page 5

Apply: Low Input

Adjust to: 50

 

Apply: High Input

Adjust to: 150

Any new settings

are canceled and

previous settings

are retained

Err. Indicates

unsuccessful

calibration

Goes directly

to Zero setting

if Analog Output

is not installed

 TWO POINT DIGITAL

CALIBRATION MODE

See Page 4

+

4.00

+

20.00

1  Dimmest

2  Dim        

3  Bright     

4  Brightest

SETPOINT SETTING  AND 

RELAY CONFIGURATION MODE

See Page 8

L or h

L or h

L or h

L or h

Display

Brightness

            [br]

Pump On Pump Off

When PUM is selected ON, 

and SP2 is set at a value 

higher than SP1, the SP1 

relay will operate in 

a special "pump on pump off" 

Hysteresis mode.  SP2 acts 

as the upper limit and SP1 

acts as the lower limit of the 

Hysteresis Band.

For filling applications, rLYS 

should be set to LhLh.  SP1 will 

then activate for inputs less than 

the SP1 setpoint, and remain 

ON until the SP2 setpoint is 

reached.

For emptying applications

rLYS should be set to hhhh.  

SP1 will then activate for inputs 

greater than the SP2 setpoint, 

and remain ON until the SP1 

setpoint is reached.

Note: The functions in grayded area 

are not applicable for this model.

Software Version is Displayed on Power-up

When power is applied, all segments of the bargraph and dig-

ital display light up for 3 seconds.  The version number of the 

installed software is then displayed for 2 seconds, after which, 

the operational display indicates the input signal.

15 Second Program Timeout

Except for ZERO and SPAN settings in the Two Point Digital Calibration Mode 

and the Analog Output Range Setting and Calibration Mode (cLo and chi), the 

meter  has  a  15  second  program  timeout.  If  no  buttons  are  pressed  for  15 

seconds in any of the other programming sequences, the meter will exit the 

programming mode and return to the operational display. Any program chang-

es that were made prior to pressing the 

P

 button in the preceding step will 

not be saved.

Содержание Leopard Series

Страница 1: ...or Bargraph Indicator Bargraph center zero function Digital display blanking Four level brightness control accessed by the button and adjusted by the button Software Features Bargraph Center Point Display Mode Selection 5 Case Dimensions 9 Component Layout 7 Connector Pinouts 6 Connectors 6 Controls and Indicators 2 Decimal Point and Brightness Selection 5 Software Features 1 Software Logic Tree 3...

Страница 2: ...at an alternative logic branch will be followed when a particular option is present When two buttons are shown side by side and enclosed by a dotted line they must be pressed at the same time then released to go onto the next programming step If an X appears through a digit it means that any number displayed in that digit is not rel evant to the function being explained P Span 10000 P P When the a...

Страница 3: ...ut Lo cLo will occur This branch will only appear if the analog output option is installed DIGITAL SPAN SELECTION FOR ANALOG RANGE OUTPUT See Page 6 DIGITAL DISPLAY ON OFF See Page 5 BARGRAPH CENTER POINT DISPLAY MODE SELECTION See Page 6 TWO POINT ANALOG OUTPUT RANGE SETTING AND CALIBRATION See Page 7 Set the Decimal Point dp Select XX XX Bargraph Display Scale Lo Range Setting bLo Adjust to 50 B...

Страница 4: ...Digital Span Selection for Bargraph Display and the display flashes bhi and the previous setting ERROR Indicates Unsuccessful Calibration If the calibration was unsuccessful the display indicates Err the new calibration settings just entered will not take effect and the previously stored setting will remain The three most likely causes of an error during calibration are 1 The full scale and zero s...

Страница 5: ...ous brightness setting STEP D Set the Bargraph and Digital Display Brightness 1 Using the and buttons adjust the display to the desired brightness setting 4 is the brightest setting 2 Press the P button Display toggles between Cto and oFF Digital Span Selection For Bargraph Display Decimal Point and Brightness Selection Bargraph Center Point Display Mode Selection STEP G Bargraph Center Point Mode...

Страница 6: ...a unique plug and socket outline to prevent cross connection The main board uses standard right angled connectors Replacement 2 3 and 4 pin plug connectors are available see Accessories on page 10 Input Signal Pins 1 to 6 Pins 1 to 6 are reserved for the input signal conditioner See the data sheet for the selected input signal conditioner Pins 17 to 21 Rear Panel Switches Pin 19 Programming LOCK B...

Страница 7: ...ODULE 24V Exc IP07 Universal Process Input 2V 5V 10V 20V 200V 2mA 20mA Custom Custom 20mA 2mA 200V 20V 10V 5V 2V VOLTAGE CURRENT UNIVERSALPROCESS Processinput Requires Digital Calibration HI LO OFF ON 24VEXC 224C Input Range Header IP07 24V Exc on off Header Signal Type Header Voltage Current ...

Страница 8: ...l adapter enables modern DIN meters to fit in existing cutouts individually or stacked when replacing old 6 edgewise mechanical pointer meters Fits existing cut outs for 6 150 mm edgewise switchboard pointer meters from Crompton G E Westinghouse Yokogawa and most others Width 43 7 mm to 48 mm 1 72 to 1 89 Height 143 4 mm to 149 mm 5 62 to 5 88 Fits 6 Edgewise Pointer Meter Cut Outs Panel Adapter W...

Страница 9: ... IO board The adapter snaps on the 36x144 mm 1 42 x5 69 case and enables single unit or stack mounting in an existing 6 edgewise pointer meter cut out 100 0 25 50 75 100 0 25 50 75 SP1 SP2 SP3 SP4 Zero Span Zero Span Panel adaptor plates are available to retrofit most existing panel cutouts 31mm 1 22 These dimensions are increased by 2mm 0 08 when the metal surround case is installed For extra str...

Страница 10: ...odel number FL B101D40PF 144x36mm 101 Segment Bargraph 4 Digit Scale Factor Standard Options for this Model Number Order Code Suffix Description Display VRR Red LED Bargraph w 4 Digit Red DPM Vertical VGG Green LED Bargraph w 4 Digit Green DPM Vertical VGR Green LED Bargraph w 4 Digit Red DPM Vertical VRG Red LED Bargraph w 4 Digit Green DPM Vertical HRR Red LED Bargraph w 4 Digit Red DPM Horizont...

Отзывы: