46
8-7-16. Sending all preset values (Response to the ST1 command)
MS1
,**,****,****,**・・・**,****
・
Data of all preset memories requested by the ST1 command are sent in integer.
MS1
,
**
,
*
*
*
*
,
*
*
*
*
,
*
*・・・*
*
,
*
*
*
*
Data of the preset memories are
sent in sequence.
System address
※
Preset data are sent in integer in the following order:
PRESET 4 set voltage, PRESET 4 set current, PRESET 1 set voltage, PRESET 1 set current,
PRESET 2 set voltage, PRESET 2 set current, PRESET 3 set voltage, PRESET 3 set current
8-7-17. Sending all equipment statuses (Response to the ST2 command)
MS2
,**,
1
,*,*
000
,*
・
All statuses of the units requested by the ST2 command are sent.
The parameters are sent in the order shown below.
MS2,
**
,
Preset currently selected
(0: Preset 4
1: Preset 1
2: Preset 2
3: Preset 3)
OUTPUT SELECT key ON/OFF (0: OFF 1: ON)
1
* *
0 0 0, 0
MAIN OUTPUT ON/OFF (0: OFF 1:ON)
System address
8-7-18. Sending ID data (Response to the ST3 command)
MS3
,**,**
・
The system address and equipment ID are sent in response to the ST3 command.
MS3,
**
,
**
Equipment ID is displayed
15: PAR18-6A 16: PAR36-3A
System address
8-7-19. Sending output voltage, output current or operation mode (Response to the ST4 command)
・
The output voltage, output current, and operation status requested by the ST4 command are sent
in real numbers. The message format is the same as MS0 shown in "8-7-15.
Sending output
voltage, output current or operation mode (Response to the ST0 command)".
8-7-20. Sending all preset values (Response to the ST5 command)
・
Data in all preset memories requested by the ST5 command are sent in real numbers.
The message format is the same as MS1 shown in "8-7-16.
Sending all preset values
(Response to the ST1 command)
".