Quick Start Guide
www.ti.com
2.2.2
ADS54J60EVM Setup
Next, setup the ADS54J60EVM using the following:
1. Connect the included 5-V power supply cable to connector J9 of the EVM. Connect the red wire to
+5VDC +/- 0.1VDC of a power supply rated for at least 3 Amps. Connect the black wire to GND of the
power supply.
2. Connect the included mini-USB cable to the USB connector J8.
3. Turn on the power supply. The power draw should be around 0.66 Amps. When the board is
configured, it will draw approximately 1.35 Amps.
4. Set the analog input signal generator for 170 MHz, and about +15 dBm of power.
5. Place a narrow pass-band band-pass filter at the output of the analog signal generator to remove noise
and harmonics from the signal generator.
6. Connect the analog input signal generator to the EVM though SMA connector AINP (J2).
2.3
Software Setup Procedure
The software can be opened and configured once the hardware is properly setup.
2.3.1
ADS54J60 GUI Configuration
1. Open the ADS54Jxx EVM GUI by going to
Start Menu
→
All Programs
→
Texas Instruments ADCs
→
ADS54Jxx EVM GUI
.
2. Verify that the green
USB Status
indicator is lit in the top right corner of the GUI. If it is not lit, click the
Reconnect USB
button and check the
USB Status
indicator again. If it is still not lit, then verify the EVM
is connected to the computer through the included mini-USB cable.
3. Click on the
Low Level View
tab then click the
Load Config
button.
4. Navigate to
C:\Program Files(86)\Texas Instruments\ADS54Jxx EVM GUI\Configuration Files
, select
the file called
LMK_Config_Onboard_983p04_MSPS.cfg
, then click
OK
. This programs the LMK04828
to provide a 983.04 MHz clock to the ADC.
5. Verify that the LMK04828 phase lock loop (PLL) is locked by checking that the
PLL2 LOCKED
LED
(D3) is lit.
6. Once the LMK04828 PLL is locked, press SW1 (
ADC RESET
) to provide a hardware reset to the ADC.
This switch is located in the middle of the EVM.
7. In the
Low Level View
tab, click
Load Config
. Select the file called
ADS54J60_LMF_8224.cfg
and click
OK
. The ADS54J60EVM is now configured for no decimation and 8 JESD204B lanes.
6
ADS54J60EVM
SLAU629A – April 2015 – Revised January 2016
Submit Documentation Feedback