170 MHz, +15 dBm
Low-Noise
Power Supply
5 VDC @ 3 A
Power
Switch
TSW14J56EVM
PC Running
HSDC Pro
Mini-USB Cable
170-MHz
Bandpass
Filter
USB 3.0 Cable
ADS54J60
EVM
AINP
BINP
Analog Input
Low-Noise
Signal Generator
5-VDC Power Supply Cable
Low-Noise
Power Supply
5 VDC @ 3 A
www.ti.com
Quick Start Guide
2.2
Hardware Setup Procedure
A typical test setup using the ADS54J60EVM and TSW14J56EVM is shown in
Figure 2
. This is the test
setup used for the quick start procedure. The rest of this section describes the hardware setup steps.
Figure 2. Quick Start Test Setup
2.2.1
TSW14J56EVM Setup
First, setup the TSW14J56EVM using the following steps:
1. Connect the ADS54J60EVM to the TSW14J56EVM using the FMC connectors.
2. Connect the included power supply cable to connector J11 (+5V IN) and the other end to a +5VDC +/-
0.3VDC 3 Amp power supply.
3. Connect the included mini-USB cable to the USB connector (J9).
4. turn on the power supply. Flip the power switch (SW6) to the
ON
position. The board should draw
around 0.5 Amps after power up. This will increase to around 1.7 Amps when loaded with firmware.
5
SLAU629A – April 2015 – Revised January 2016
ADS54J60EVM
Submit Documentation Feedback
Copyright © 2015–2016, Texas Instruments Incorporated