![Texas Instruments ADS1298RECG-FE Скачать руководство пользователя страница 47](http://html1.mh-extra.com/html/texas-instruments/ads1298recg-fe/ads1298recg-fe_user-manual_1095391047.webp)
ADS1298R Daughter Card Hardware Details
47
SBAU181B – March 2011 – Revised Janurary 2016
Copyright © 2011–2016, Texas Instruments Incorporated
ADS1298R
Table 5. ADS1298RECG-FE Default Jumper/Switch Configuration (continued)
Jumper
Function
Settings
(2)
Requires installation of U2
(3)
Requires installation of JP3 and references U3/U4
JP17
ECG shield drive buffer
input(Header Not Installed)
(2)
1-2: ECG shield drive connected to
RLDOUT
2-3: ECG shield drive connected to
RLDINV
JP18
CLK
connection
1-2:
CLK
connected to DSP (J3.17)
2-3:
CLK
connected to OSC1
JP19
OSC1 Enable
1-2: OSC1 enabled
2-3: OSC1 disabled
JP20
AVSS supply source
1-2: AVSS selected for single supply operation (AVSS = 0V (AGND))
2-3: AVSS selected for bipolar supply operation (AVSS = -2.5V)
JP21
CS
source
1-2:
CS
connected to DSP via J3.1
2-3:
CS
connected to DSP via J3.7
JP22
START
source
1-2:
START
comes from J3.1
2-3:
START
comes from J3.14
JP23
CLKSEL
source
1-2: External Master Clock
2-3: Internal Master Clock (
CLKSEL
controlled by J3.2 (pulled up to DVDD))
JP24
DVDD Supply Select
1-2: DVDD supply = 1.8V
2-3: DVDD supply = 3.3V
JP25
External Reference Selection
(Header Not Installed)
(3)
1-2: Select U4 as reference source
2-3: Selected U3 as reference source
JP26
1-2: CH8- connection
Open: Channel input not connected
Installed: Channel input connected to
WCT
(requires JP16 to be installed)
3-4: CH8+ connection
Open: Channel input not connected
Installed: Channel input connected to ECG_V1
JP27
1-2: CH7- connection
Open: Channel input not connected
Installed: Channel input connected to
WCT
(requires JP16 to be installed)
3-4: CH7+ connection
Open: Channel input not connected
Installed: Channel input connected to ECG_V5
JP28
1-2: CH6- connection
Open: Channel input not connected
Installed: Channel input connected to
WCT
(requires JP16 to be installed)
3-4: CH6+ connection
Open: Channel input not connected
Installed: Channel input connected to ECG_V4
JP29
1-2: CH5- connection
Open: Channel input not connected
Installed: Channel input connected to
WCT
(requires JP16 to be installed)
3-4: CH5+ connection
Open: Channel input not connected
Installed: Channel input connected to ECG_V3
JP30
1-2: CH4- connection
Open: Channel input not connected
Installed: Channel input connected to
WCT
(requires JP16 to be installed)
3-4: CH4+ connection
Open: Channel input not connected
Installed: Channel input connected to ECG_V2
JP31
1-2: CH3- connection
Open: Channel input not connected
Installed: Channel input is connected to ECG_RA
3-4: CH3+ connection
Open: Channel input not connected
Installed: Channel input is connected to ECG_LL
JP32
1-2: CH2- connection
Open: Channel input not connected
Installed: Channel input is connected to ECG_RA
3-4: CH2+ connection
Open: Channel input not connected
Installed: Channel input is connected to ECG_ LA
JP33
1-2: CH1- connection (Chest
Lead 6 connected)
Open: Channel input not connected
Installed: Channel input connected to
WCT
(requires JP16 to be installed)
3-4: CH1+ connection (Chest
Lead 6 connected)
Open: Channel input not connected
Installed: Channel input connected to ECG_V6
5-6: CH1- connection
(Respiration connected)
Open: Channel input not connected
Installed: Channel input connected to CH1- input signal mux (see JP34)
7-8: CH1+ connection
(Respiration connected)
Open: Channel input not connected
Installed: Channel input connected to CH1- input signal mux (see JP35)