Using the ADS1298RECG-FE Software
30
SBAU181B – March 2011 – Revised Janurary 2016
Copyright © 2011–2016, Texas Instruments Incorporated
ADS1298R
3.5.4
RESP Tab
The Resp tool allows the user to make use of the on-board respiration circuitry to show a simulated
respiration signal.
shows the RESP tab tool features.
Respiration Tab
The bottom left portion of the screen provides controls that may be utilized to add filtering to the displayed
respiration signal. NOTE: The filtering performed using this analysis tool is performed in the UI software
and is not available in the ADS1298R device. The
Low Pass Filter
group of controls implement a low-
pass digital filter on the collected data. The filter is enabled using the
Enable
control, and
Filter Order
and
Cutoff Frequency
allow for the customization of the filter. The
High Pass Filter
group of controls
implement a digital high-pass filter on the collected data. The high-pass controls function the same as the
low-pass controls except that they apply to the high-pass filter. The
Unsettled Points to Remove
control
allows the user to remove a specific number of points from the beginning of the filtered data set. This
allows the user to remove the data the isn't filtered and look only at the filtered data.
The
Simulated Respiration Frequency
set the respiration frequency that the on-board respiration
circuitry will simulate when collecting data.
The
Statistical Data
group provides two calculated parameters from the displayed data,
Mean
and
Noise
.