Operation
Testo Sensor LD basic V1.01 EN
Seite
9
von
22
10
Start-up / / Application Testo Sensor LD basic
Please first observe the safety instructions in Chapter 2
10.1
Switch on
Hold down the power button for about 1 second, the power will turn on, and a start-up sequence
will appear on the display. Pressing the button again switches the device off again.
device components and controls
10.2
Headphone Volume Up / Volume Down
The volume up and volume down buttons in the headset can be increased or decreased in 16
steps.
Continuously pressing the button automatically increases / decreases the value.
Volume up / down buttons for headphone volume, see
device components and controls
Please make sure the headphone level is <50% before putting on the headphones.
10.3
Sensitivity level
Ultrasound levels can be understood as a "loudness" of the leakage.
With the “Sensitivity" button, the sensitivity of the
Testo Sensor LD basic can be adjusted to the
environment, which strongly influences the acoustic behaviour of the device and increases or
decreases the valid value range. A reduction in sensitivity reduces the range of the leakage .
The sensitivity levels or the automatic function are selected via the "Sensitivity" button.
Sensitivity levels
0
–
60 dB
= Highest sensitivity level of the device (use with small leaks and no noise
10
–
70 dB
= Leakages and noises get "less noisy", the range is reduced.
20
–
80 dB
= Leakages and noises get "less noisy", the range is reduced.
30
–
90 dB
= Leakages and noises get "less noisy", the range is reduced.
40
–
100 dB
= Most insensitive stage (large leaks, many noises
→
for heavy-duty application)
The LD basicoffers a manual selection as
well as an automatic function
"Auto
" for
setting the sensitivity.
Selection is made by pressing the
"Sensitivity"
setting button.
When
"Auto"
is selected, the LD basic
selects the best possible sensitivity level.
Sensitivity
Button