8 Technical data
99
Feature
Value
Protection class
IP68
Conformity with standards
DIN EN 12830
Operating temperature
-35 to +50°C
8.2.3
testo Saveris T2 / T2D
Saveris radio data logger with external probe connection and internal NTC, door
contact.
Feature
Value
Probe type (internal)
NTC
Measuring range (internal)
-35 to +50°C
Accuracy (internal)
±0.4°C (-25 to +50°C)
± 0.8°C (remaining measuring range)
Resolution (internal)
0.1°C
Probe type (external)
NTC
Measuring range (external)
-50 to +150°C
Accuracy (external)
±0.2°C (-25 to +70°C)
± 0.4°C (remaining measuring range)
Resolution (external)
0.1°C
Connection
NTC via mini-DIN socket, door contact
connection cable included in the scope of
delivery (1.80 m). The setting of the door
contact switch affects the service life.
Protection class
IP68 (with probe inserted or connection socket
sealed with rubber plug)
Conformity with standards
DIN EN 12830
Operating temperature
-35 to +50°C