4.2 Accessories and spare parts

  147 

 

Calibration 

Standard ISO calibration certificate, trans probes 

0520 0176 

Special ISO calibration certificate, trans probes 

0520 0066 

Standard DAkkS calibration certificate, trans probes 

0520 0276 

Special DAkkS calibration certificate, trans probes 

0520 0236 

ISO- calibration certificate humidity, probes 

0520 0076 

DAkkS calibration certificate temperature, probes 

0520 0261 

 

Содержание 6681 Profibus

Страница 1: ...testo 6681 Humidity transmitters testo 6610 Probes P2A software Parameterizing adjusting and analyzing software Instruction manual Volume 1 en...

Страница 2: ...Observe the permissible storage transport and operating temperatures Never store or operate the product with solvents and do not use any desiccants Do not use the instrument for control purposes at th...

Страница 3: ...ards which could result in serious physical injury if the precautionary measures indicated are not taken e g De energize the mains connection before connecting the transmitter With the signal word Cau...

Страница 4: ...6681 until it engages to Example entries are in inverted commas e g The value 0 results in Bold type Elements of the program interface or instrument display e g The instrument designation appears in...

Страница 5: ...3 1 Assembling the instrument 16 1 3 2 Connecting the instrument 19 1 3 3 Adjusting the instrument 31 1 4 Operation 41 1 4 1 Relationship between user menu and mini DIN socket is active 41 1 4 2 Key c...

Страница 6: ...information about commissioning operating and maintaining the testo 6610 probe The testo 6681 humidity transmitter is suitable for the following applications for example Process instrumentation Clean...

Страница 7: ...be found in volume 2 chapter 4 2 or on the website at www testo com 1 1 4 Technical Data Parameters Humidity various variables and units Temperature C F Measuring range Depends on probe Accuracy Depe...

Страница 8: ...plain text line optional Housing operating temperature 40 to 70 C 40 to 158 F with display from 0 50 C 32 to 122 F With integrated relay 40 to 60 C Storage temperature 40 to 80 C 40 to 176 F Housing...

Страница 9: ...ction e g voltage supply 7 Earthing PE connection 8 M 20 x 1 5 screw connection e g relay R 3 and R 4 9 Eyelet for measuring point panel 10 M 20 x 1 5 screw connection e g relay R 1 and R 2 11 Probe c...

Страница 10: ...able with cover electrode monitoring probe version temperature range 40 C to 180 C 40 to 356 F 1 2 3 Display and keypad The display option allows the testo 6681 humidity transmitter to be operated via...

Страница 11: ...81 has either 2 or optionally 3 current outputs of 4 to 20 mA 2 wire 0 to 20 mA 4 wire 4 to 20 mA 4 wire or 2 or optionally 3 voltage outputs of 0 to 1 V 0 to 5 V 0 to 10 V 4 wire The transmitter can...

Страница 12: ...sation appears as of a displayed humidity starting from 70 and is shown on the display This unit is used in meteorology for example According to the WMO the Magnus coefficient is used for sub cooled w...

Страница 13: ...0 of X Min value of standard 50 of X It is thus possible to scale beyond the measuring range e g for the adjustment of the scaling limits to standard values of a PLC However the physical measurement r...

Страница 14: ...d 6615 60 30 80 100 Ftd 6615 148 212 112 212 Mixture dewpoint H2O2 Ctm 20 100 20 100 Ftm 4 212 4 212 Absolute humidity g m3 all probes 0 600 0 2000 gr ft3 0 250 0 800 Degree of humidity g kg all probe...

Страница 15: ...is switched If the reading reverts to more than a specified hysteresis below or above the limit value the alarm is cancelled In addition information about the occurrence of error status messages can b...

Страница 16: ...panel bracket 1 Remove locking screw item 4 of drawing below and detach rear panel bracket from plastic bracket item 2 of drawing below 2 Hold rear panel bracket in assembly position and mark the thre...

Страница 17: ...t to rear panel bracket 1 Slide plastic bracket 2 on the back of instrument onto rear panel bracket until it engages see arrows 2 Insert screw 4 through hole 3 and screw into rear panel bracket 3 Inse...

Страница 18: ...probe shaft 2 Drill a hole 12 5 mm in the duct wall to feed through the probe shaft 3 Fasten wall duct bracket 6 to duct wall with screws 5 4 Push probe shaft 9 with filter 10 through the middle hole...

Страница 19: ...r onto bracket 3 4 until it engages Take the weight of the transmitter into account Ensure that the brackets 4 6 are fastened securely 7 Insert screw 1 through the hole on the top of the instrument an...

Страница 20: ...pper part of housing from lower part 3 and place on a clean surface Warning Electrical voltage Danger of injury De energize the mains connection before connecting the transmitter Only have the transmi...

Страница 21: ...erminal strip for voltage supply and analog outputs 6 Terminal board 7 Earthing terminal internal 8 M 16 x 1 5 screw connection 9 Earthing terminal external 10 M 20 x 1 5 screw connection 11 Eyelet fo...

Страница 22: ...ge supply and analog signal lines through opened M 16 x 1 5 screw connection item 8 in Overview of terminals chapter 1 3 2 1 2 Strip the cable ends clamp wire end ferrules on and screw down onto volta...

Страница 23: ...st exceeding 8 A An OFF switch must be installed in an easily accessible position close by and be marked as such 1 Feed connection cables of the two or optionally three channels through opened M 16 x...

Страница 24: ...screw connection 1 Feed connection cables for the relays through opened M 20 x 1 5 screw connection item 10 of Overview of terminals chapter 1 3 2 1 2 Strip cable ends and clamp on wire end ferrules 3...

Страница 25: ...cable sheathed cable with a cross section of at least 1 5 sq mm must be used Cable connection 2 may not be routed in a loop within the tray 1 It is recommended that you always tie 3 cores to one anoth...

Страница 26: ...s or the circuit is interrupted This circuit can therefore be used to monitor the functionality of the alarm circuit as a cable break for instance is indicated by the busy light going off Use of relay...

Страница 27: ...ing operation 5 Close M 20 x 1 5 screw connection item 10 in Overview of terminals chapter 1 3 2 1 1 3 2 4 Plug in connection option As an option the PG screw connections of the signal and supply line...

Страница 28: ...supply and channels M12 plug in connection 5 pin socket 1 View of the plug in connections in the installed state from outside Pin Assignment 1 V 24 2 V 24 3 Ch 1 4 Ch 1 5 PE M12 plug in connection 5 p...

Страница 29: ...n the instrument 1 Guide PE line yellow green 5 through the cable coupling x and fit cable lug 8 Fix this to the side of the instrument 6 using M 5 screw 3 washer 4 and snap ring 7 on the internal ear...

Страница 30: ...to 6681 1 3 Commissioning 1 3 2 6 Closing the instrument 1 Place upper part of instrument on top of lower part see arrow and fix in place with housing screws 1 2 Close the service flap and tighten scr...

Страница 31: ...Adjustment 11 3 RH and 75 3 RH via Adjustment keys 1 2 P2A software Adjustment 20 rH and 80 rH via User menu Adjustment using a precise multimeter and transmission of analog reference value in P2A so...

Страница 32: ...in the working point The reference condition can be measured using a precise handheld instrument or be created in an air conditioning cabinet The advantage of the 1 point adjustment is the good measur...

Страница 33: ...justed to the reference value at the two standard adjustment points 11 3 RH and 75 3 RH or 20 RH and 80 RH The reference conditions are created either by using Testo humidity adjustment set order no 0...

Страница 34: ...out via the user menu A 2 point adjustment cancels any previous 1 point adjustment Adjustment with the standard Testo adjustment salt pots is not suitable for the testo 6614 heated for high humidity a...

Страница 35: ...for the adjustment points 20 RH and 80 RH is carried out via the user menu The service flap of the testo 6681 is open 1 Expose the humidity probe of the testo 6681 to the reference condition of 11 3...

Страница 36: ...eset status message is shown Carry out the adjustment analogously for the reference condition 75 3 RH Press on the 75 3 RH adjustment key 6 to do this 3 Close the service flap 1 3 3 4 Analog output ad...

Страница 37: ...e testo 6681 and close service flap Adjusting analog output 3 optional If the optional third analog output is to be adjusted a cable connection to measure the analog value must be installed Do this by...

Страница 38: ...ondensation range and the risk of the sensor corroding is also reduced Using the separate temperature probe the testo 6681 transmitter compensates the microclimate conditions and displays the process...

Страница 39: ...s This means that extremely accurate measuring results are also attained to 60 Ctd To this end a temperature sensor is fitted on the back of the testo 6615 which is used as a heater A humidity and tem...

Страница 40: ...t vary excessively If these prerequisites cannot be fulfilled the values obtained during the last successful self adjustment will be maintained All adjustment processes are saved in the history of the...

Страница 41: ...In the factory setting a third adjustment point 40 Ctd is approached for the testo 6615 in addition to the 2 point adjustment This special adjustment can be performed again by your Testo Service team...

Страница 42: ...the key cover has been assembled the service flap must be opened for operation see Section Opening the instrument chapter 1 3 2 Attaching the key cover The service flap is open see Opening the instru...

Страница 43: ...to be used the numerical code 0000 must be entered This is also the status upon delivery 1 4 4 Structure of user menu At the main menu level the user menu comprises the following Main menu of channel...

Страница 44: ...Mode changes to parameterization In Parameterizing Mode confirms a selection or setting ESC Leaves a menu without modifying any settings Selecting scrolls through menus downwards or selectable alterna...

Страница 45: ...testo 6681 1 4 Operation 45 1 4 5 Overview of the testo 6681 user menu...

Страница 46: ...46 testo 6681 1 4 Operation...

Страница 47: ...l 1 The lower scale limit is edited Unit as selected above example 4 mA 0 RH Editing the value Scroll one digit to the right using increase value of digit by 1 using Confirm with SET or abort entry wi...

Страница 48: ...alarm 1 In Measuring Mode press SET select Main Menu Alarm using or and confirm selection with SET Four alarms can be parameterized 2 Select Alarm x using or and confirm selection with SET Using alarm...

Страница 49: ...chapter 3 Alarm is selected see previous steps 1 and 2 1 Specify with or whether Alarm x is to be used as the Alarm relay or is not to be used Confirm selection with SET and return to Alarm x 2 Chang...

Страница 50: ...t Contrast The brightness difference between the display background and the displayed values is changed Edit select parameter with or confirm with SET or abort input with ESC the effect of the change...

Страница 51: ...uid end solution Scroll one digit to the right using increase value of digit by 1 using Confirm with SET or abort entry with ESC 4 Go back to H2O2 by pressing ESC and continue to Absolute pressure uni...

Страница 52: ...of analog and relay outputs In addition you can read off the minimum and maximum readings since the last voltage supply or reset of the min max values Testing functionality of analog outputs This fun...

Страница 53: ...F and ON using or If ON is chosen the NO contact is closed the NC contact opened If OFF is chosen the NC contact is closed the NO contact opened 3 To test route a measuring cable from the relay termin...

Страница 54: ...with SET 4 Scroll between the messages recorded so far using or and press ESC to return to Last messages 5 Continue to Display of message with or ON Measurements are shown on the display in Measuring...

Страница 55: ...using or or return to Measuring Mode by pressing ESC 1 4 6 9 Editing Adjust main menu A reference value can be entered for both relative humidity RH and temperature C F for the 1 point adjustment Plea...

Страница 56: ...ith or 7 Press SET edit value Scroll one digit to the right using increase value of digit by 1 using Confirm with SET or abort entry with ESC 8 Continue to Analog Adj Ch 1 with or 9 Continue with the...

Страница 57: ...onfirm with SET or abort entry with ESC 5 Select Adj Point 2 using or 6 Press SET Read off multimeter display e g 12 001 mA and enter this value in the user menu Do this by scrolling one digit to the...

Страница 58: ...at the time of supply to the customer 1 In Measuring Mode press SET select Main Menu Reset using or and confirm selection with SET Reset device to factory settings is displayed 2 Select the setting t...

Страница 59: ...transmitter the last 180 messages are stored in a ring memory but there is no restriction in the P2A software 1 5 1 Status messages Status messages show the current operating mode of the testo 6681 M...

Страница 60: ...n automatic self adjustment 02518 Probe reset Probe reset The probe performs a reset 1 5 2 Warning messages Warning messages show an early warning or a current malfunction which may negatively impact...

Страница 61: ...ation developing Take measures to reduce process humidity 02807 Values less than 0 RH The adjustment or sensor is faulty Check adjustment via P2A adjustment history perform 2 point adjustment where ne...

Страница 62: ...tch the 665x transmitter and the 661x probes the 668x transmitter 01528 Watchdog error Due to a processor error the transmitter performs an automatic restart If the problem occurs frequently contact T...

Страница 63: ...x Transmitter refresh x Probe connection Service plug User Setting Change x Probe disconnected Analog out adjust x x 1 point adjustment x x 2 point adjustment 11 3 x x 2 point adjustment 75 3 x x Prob...

Страница 64: ...ction which means that the collective alarm is activated as soon as at least one of the messages assigned to it is activated The collective alarm can be assigned to each of the 4 optional relays The c...

Страница 65: ...Display message Class 0 20mA 4 20mA 1V 5V 10V No probe signal Error 21mA 21mA 1 2V 5 5V 11V Wrong probe Error 21mA 21mA 1 2V 5 5V 11V Watchdog error Error 21mA 21mA 1 2V 5 5V 11V Values 0 RH Underran...

Страница 66: ...4 or P2A software volume 2 chapter 3 Convenient remote monitoring of the transmitter can be implemented for example by using a relay as a collective alarm see chapter 1 4 6 4 whose messages are forwar...

Страница 67: ......

Страница 68: ...Testo SE Co KGaA Celsiusstr 2 79822 Titisee Neustadt Germany Tel 49 7653 681 0 E Mail info testo de www testo com 0970 6681 en 05 Vol1...

Страница 69: ...testo 6681 Humidity transmitters testo 6610 Probes P2A software Parameterizing adjusting and analyzing software Instruction manual Volume 2 en...

Страница 70: ......

Страница 71: ...3 2 3 2 Connecting removing the probe to from the transmitter 107 2 4 Maintenance and cleaning 107 2 4 1 Replacing filters protection caps 107 2 4 2 Cleaning the instrument and filter protection cap 1...

Страница 72: ...TIPS AND ASSISTANCE 144 4 1 Questions and answers 144 4 2 Accessories and spare parts 145 4 2 1 Ordering options for testo 6681 transmitter 0555 6681 148 4 2 2 Ordering options for testo 6610 probes...

Страница 73: ...d transmit their adjustment data to the internal memory of the testo 6681 transmitter The information is transmitted between the probe and transmitter in a purely digital form The probes can therefore...

Страница 74: ...the perfect way the individual layers are matched to each other This is evident particularly in the top electrode which has to perform two tasks that at first glance appear contradictory it must be c...

Страница 75: ...y issued once a clear effect is discernible Early warning when sensor corrosion is starting to develop The testo 6617 probe is able to signal the first signs of corrosion The probe can thus be changed...

Страница 76: ...escription of the probe versions can be found from chapter 2 2 2 onwards The following probe versions are available for the testo 6681 humidity transmitter Version Article no Characteristic testo 6611...

Страница 77: ...C 40 to 356 F 2 2 1 2 Determining the accuracy measuring uncertainty The entries for the measuring uncertainty for the probe are determined in accordance with GUM Guide to the Expression of Uncertain...

Страница 78: ...or the parameter once coming from a smaller value once coming from a larger value humidity sensors actually have no hysteresis but rather very slow adjustment effects that appear to be hysteresis when...

Страница 79: ...er with drip hole and condensation protector M 08 Filter for H2O2 atmospheres Nxx Cable length N 00 Without cable testo 6611 N 01 1 m cable length testo 6613 6614 6615 6617 N 02 2 m cable length testo...

Страница 80: ...554 0759 Sintered PTFE filter 35 M 04 0554 0755 Open metal protection cap 35 M 05 0192 0265 Open ABS plastic cap 25 M 06 0554 9913 PTFE filter with drip hole 35 M 07 0554 9913 0554 0166 PTFE filter wi...

Страница 81: ...wired At a glance Application Monitoring and regulating the production and storage air quality when manufacturing and storing hygroscopic products Meeting the top accuracy requirements Clean room appl...

Страница 82: ...erature 0 15 C 0 27 F with slope PT1000 class AA Refer to the charts below for the correlation between temperature and accuracy Length 70 mm As with length of 200 mm but with additional measuring erro...

Страница 83: ...wall probe Humidity error according to amount RH as a factor of process humidity Temperature error as a factor of process temperature and temperature of electronics RH RH System error 6681 probe elect...

Страница 84: ...s At a glance In the event of overpressures the probe may become a projectile For assembly see Pressure resistance on the following page 1 Filter including humidity and temperature sensor 2 Probe shaf...

Страница 85: ...30 to 150 C 22 to 302 F Material Probe shaft Stainless steel Line Sheathed FEP Connector ABS plastic Accuracy at 25 C 77 F Humidity 2 For constant use in high humidity 80 rF at 30 C for 12 h 60 rF at...

Страница 86: ...300 500 800 mm L A 165 265 465 765 mm A see Table Filters chapter 2 2 1 4 Cable length incl probe shaft and filter Customized for duct version Pressure resistance PN 10 probe tip If installing probe...

Страница 87: ...Product description 87 Temperature error as a factor of process temperature and temperature of electronics System error 6681 probe electronics 25 C 77 F System error 6681 probe electronics 25 C to 70...

Страница 88: ...e following page Application Monitoring and regulating industrial humidity processes apart from high humidity processes e g food production storing fruit Monitoring the production and storage air qual...

Страница 89: ...tic Accuracy at 25 C 77 F Humidity 1 0 RH 0 007 x reading for 0 to 90 RH 3 For constant use in high humidity 80 rF at 30 C for 12 h 60 rF at 30 C for 12 h please contact us via www testo com service c...

Страница 90: ...5 mm A see Table Filters chapter 2 2 1 4 Probe length incl probe shaft and filter 1 2 5 10 m Pressure resistance PN 10 probe tip PN 1 if end of probe cable is involved in process If installing probe u...

Страница 91: ...Product description 91 Temperature error as a factor of process temperature and temperature of electronics System error 6681 probe electronics 25 C 77 F System error 6681 probe electronics 25 C to 70...

Страница 92: ...lso see volume 1 chapter 1 3 3 5 for the functional principle of the testo 6614 At a glance In the event of overpressures the probe may become a projectile For assembly see Pressure resistance on the...

Страница 93: ...mperature Measuring range Humidity 0 to 100 RH Temperature 40 to 180 C 40 to 356 F Material Probe shaft Stainless steel Line Sheathed FEP Connector ABS plastic Accuracy at 25 C 77 F Humidity 1 0 RH 0...

Страница 94: ...ter 2 2 1 4 Probe length incl probe shaft and filter 1 2 5 10 m Pressure resistance PN 10 probe tip PN 1 if end of probe cable is involved in process If installing probe under pressure please use cutt...

Страница 95: ...nificance with low humidity values in dewpoint range Please also see volume 1 chapter 1 3 3 6 for the functional principle of the testo 6615 Only use testo 6615 with sintered PTFE filter art no 0554 0...

Страница 96: ...stance on the following page Application Monitoring and regulating trace humidity processes compressed air with adsorption or membrane driers and plastic granulate driers Used when spatial separation...

Страница 97: ...Ctd 40 Ftd 4 K at 50 Ctd 58 Ftd 6 K at 60 Ctd 76 Ftd Dependency on process temperature Deviation from 25 C 77 F 0 1 Ktf K at 40 C 25 C 40 F 77 F 0 2 Ktf K at 25 C 50 C 77 F 122 F 0 4 Ktf K at 50 C 12...

Страница 98: ...tion order no 0554 1795 Measuring accuracy of testo 6615 trace humidity cable probe Dewpoint error numerically RH dependent on the process dewpoint Temperature error as a factor of process temperature...

Страница 99: ...event of overpressures the probe may become a projectile For assembly see Pressure resistance on the following page Application Monitoring and regulating industrial humidity processes involving corro...

Страница 100: ...If this damage to the sensor is not spotted until too late high costs can be incurred due to unusable readings or machine shutdowns and waiting times until the sensor is replaced 1 Cover electrode 2 B...

Страница 101: ...dependent upon the process temperature with a deviation of 25 C 77 F 4 For constant use in high humidity 80 rF at 30 C for 12 h 60 rF at 30 C for 12 h please contact us via www testo com service conta...

Страница 102: ...tting ring screw connection order no 0554 1795 Measuring accuracy of testo 6617 cable probe self monitoring Humidity error according to amount RH as a factor of process humidity Temperature error as a...

Страница 103: ...of the testo 6612 probe can be found in volume 1 chapter 1 3 1 2 2 3 1 3 Installing testo 6613 6614 6615 6617 probes If used with these probes the testo 6681 transmitter is mounted on the wall see vo...

Страница 104: ...to the humidity probe as possible max 10 cm An appropriate assembly tool is supplied with the testo 6614 Wall duct holder Order no 0554 6651 Temperature probe testo 6614 Assembly tool supplied with w...

Страница 105: ...obe Engage transmitter in the holder on the end of the duct probe Only atmospheric processes up to approx 1 bar positive pressure Alternatively the single hole duct holder order no 0554 1793 can also...

Страница 106: ...ith the testo 6614 C Process mounting During installation ensure that the probe cannot be damaged during operation For the testo 6614 heated probe version please use the cutting ring screw connection...

Страница 107: ...ted To remove the probe the lock release button on the probe must be pressed so that this can be removed 2 4 Maintenance and cleaning 2 4 1 Replacing filters protection caps 2 4 1 1 Replacing the filt...

Страница 108: ...a tool 2 4 1 2 Replacing the filter protection cap for testo 6612 duct version Do not damage the sensors when exchanging the filter the protection cap and do not touch their surfaces Hint Mark the in...

Страница 109: ...obe shaft into duct as far as the marking and fix position with screw 9 2 4 1 3 Replacing the filter protection cap for cable versions The following descriptions applies to these probes testo 6613 tes...

Страница 110: ...olvents Unscrew cap or protection cap to clean clean with compressed air and replace Take care not to damage the sensor 2 4 3 Replacing the sensor Thanks to the probe concept digital plug in all probe...

Страница 111: ...at any time from the Testo homepage www testo com download center requires registration The software must only be bought one time even for owners of several testo transmitters 3 1 1 Functions and use...

Страница 112: ...Computer The computer must fulfil the requirements of the respective operating system The following requirements must additionally be fulfilled Interface USB 1 1 or higher Graphic resolution min 1024...

Страница 113: ...re 1 Insert CD with P2A software If the installation program does not start automatically Open Windows Explorer and start the file Setup exe on the software CD 2 Follow the directions of the installat...

Страница 114: ...active at all times USB driver is installed see chapter 3 2 1 Installing the software driver 1 Start the P2A software 2 Connect adapter supplied with the P2A software to the service interface of the...

Страница 115: ...n Shows the Windows dialogue for searching and opening files Save as Saves the parameters of an instrument or parameter file under a new name Edit Copy Copies the parameters of the marked instrument o...

Страница 116: ...olbar Shows the Windows compliant icons for editing 3 File list Icon File Explanation Instrument file Instrument file Connection to the instrument has been established Type Serial number cfm File name...

Страница 117: ...ameter file was created Connection status Green connection is active Red connection is inactive 6 Status bar Shows the current status when editing via the menu bar 3 3 2 Editing instrument parameter f...

Страница 118: ...output cannot be changed Horizontal Min max scale end points of selected unit The curve changes in accordance with the entered value of scale minimum and maximum Scale minimum maximum The endpoints of...

Страница 119: ...default values Signal delay graphic Curve changes according to the set signal delay Signal delay Time interval in stages of 1 to 15 1 no delay 15 longest delay The signal delay is added to the reacti...

Страница 120: ...ays are available optional Not used Relay is not used Hysteresis image and input options are hidden Collective alarm A relay can be used as a collective alarm detector in the event that selected messa...

Страница 121: ...is it is switched to OFF NO contact or ON NC contact The graphic display in the centre of the screen refers to the relay wiring as a NO contact ON Hysteresis To avoid switching cycles Channel Selectio...

Страница 122: ...Explanation Selection of the messages error etc that should result in a collective alarm OR linkage List field with checkboxes Selection of which messages generated in the transmitter should be signa...

Страница 123: ...ation of the following units CtdA or FtdA g kg or gr lb ppmV vol H2O2 process The calculation of the mixture dewpoint unit Ctm is dependent upon the type of the evaporation process with passive vapori...

Страница 124: ...Display lighting is permanently switched on Display lighting when button is pressed 10 seconds When a particular button on the instrument is pressed the display lights up for 10 seconds Display brigh...

Страница 125: ...ntered Adopt new password Button for confirming the new password Valid password Display of the current password Self adjustment Explanation Parameterizing of testo 6615 probe if this is used See volum...

Страница 126: ...parameter file 2 Click on File Save as in the menu bar 3 Select storage location and enter the file name 4 Click on Save The new parameter file is shown in the file list Only the parameters are saved...

Страница 127: ...re transmitted to the file You can also use the common keyboard shortcuts for copying and pasting i e CTRL C and CTRL V Parameters can also be transmitted using drag drop where you drag the icon of th...

Страница 128: ...o test for proper function see chapter 3 3 3 4 Min max values display Overview of the minimum and maximum values measured since the last reset of the transmitter see chapter 3 3 3 5 3 Click on OK or C...

Страница 129: ...rial number dialogue is opened with the Test analyze transmitter register 2 Mark channel and test values Field button Explanation Checking the analog outputs see volume 1 chapter 1 4 6 6 Current readi...

Страница 130: ...tput using a precise multimeter Deactivate Finish entering the electrical variables at the analog output The analog output returns to the Current reading again 3 Click on OK or Cancel to close the dia...

Страница 131: ...s that the value is being transmitted to a connected PLC external display etc in the event of existing cabling Deactivate relay n Open contact A warning informs that the value is being transmitted to...

Страница 132: ...ick on Test analyze transmitter The Properties of Instrument type Serial number dialogue is opened with the Test analyze transmitter register 2 Mark Min max values Field button Explanation View the mi...

Страница 133: ...entry via assistant wizard Also see volume 1 chapter 1 3 3 for this 3 3 4 1 1 point adjustment 1 Expose the reference measuring instrument and the instrument to be adjusted to the same constant condi...

Страница 134: ...ated every second Reference value Entry of the read off value from the reference measuring instrument Permissible entries max 5 RH deviation sum of all 1 point adjustments max 2K C deviation sum of al...

Страница 135: ...instrument file of the connected instrument 3 Click on Adjusting the transmitter The Properties of Instrument type Serial number dialogue is opened with the Adjusting the transmitter register 4 Mark...

Страница 136: ...eading in RH Readings are updated every second Reference value Entry of the read off value from the reference measuring instrument Permissible entries Lower adjustment point 10 3 to 12 3 RH Upper adju...

Страница 137: ...dialogue is opened with the Adjusting the transmitter register 4 Click on Start wizard and follow the instructions of the wizard The adjustment is performed when the wizard is closed Field Explanatio...

Страница 138: ...he user menu Transmitter is entered in the User field and only the operating hour is entered in the Date time field instead of operating hour date time For entries that are performed using the P2A sof...

Страница 139: ...ettings of the operating system User Name with which the user is logged into the operating system MUF transmitter if the change was performed at the instrument Operating hours date time Operating hour...

Страница 140: ...s of the operating system User Name with which the user is logged into the operating system MUF transmitter if the change was performed at the instrument Operating hours date tim e Operating hour time...

Страница 141: ...adjustment Value before the adjustment Offset to 1 point adjustment Value after the adjustment Offset 2 point adjustment Difference between target and actual value reported by instrument Channel Analo...

Страница 142: ...tion to the P2A software Date time Format of the PC time is adopted from the settings of the operating system Operating hours Operating hour at which the message appeared in the instrument User MUF tr...

Страница 143: ...re 3 3 Using the software 143 The printing job is automatically sent to the default printer for the operating system The printout can be edited using Set up printer 4 Click on OK or Cancel to close th...

Страница 144: ...the transmitter Undo adjustment A 1 point temperature humidity adjustment can be reset to the current reading using Set offset to zero The actual values before the adjustment can be read out from the...

Страница 145: ...g analyzing incl USB adapter 0554 6020 Fastenings assembly aids Wall duct bracket with M3 screw for fastening the transmitter to the probe and the probe to the wall duct 0554 6651 Pressure tight G 1 2...

Страница 146: ...nsor filters and protective caps Stainless steel sintered filter 0554 0647 Wire mesh filter 0554 0757 Sintered PTFE filter 0554 0759 Protection cap made of metal open 0554 0755 PTFE protection cap wit...

Страница 147: ...Special ISO calibration certificate transmitter probes 0520 0066 Standard DAkkS calibration certificate transmitter probes 0520 0276 Special DAkkS calibration certificate transmitter probes 0520 0236...

Страница 148: ...AC DC B03 0 to 5 V 4 wire 24 VAC DC B04 0 to 10 V 4 wire 24 VAC DC B05 0 to 20 mA 4 wire 24 VAC DC B06 4 to 20 mA 4 wire 24 VAC DC Cxx Display C00 without display C02 with display English C03 with dis...

Страница 149: ...13 kJ kg Min Max enthalpy F14 hPa Min Max water vapour partial pressure F15 inch H2O Min Max water vapour partial pressure F18 Vol min max Gxx Channel 2 Unit G01 RH Min Max G02 C Min Max G03 F Min Max...

Страница 150: ...analog output I01 RH Min Max I02 C Min Max I03 F Min Max I04 Ctd Min Max I05 Ftd Min Max I06 g kg Min Max I07 gr lb Min Max I08 g m3 Min Max I09 gr ft3 Min Max I10 ppmVol min max I11 Cwb Min Max wet...

Страница 151: ...al languages K01 German English instruction manual K02 French English instruction manual K03 Spanish English instruction manual K04 Italian English instruction manual K05 Dutch English instruction man...

Страница 152: ...PTFE filter with drip hole and condensation protector M 08 Filter for H2O2 atmospheres Nxx Cable length N 00 Without cable testo 6611 N 01 1 m cable length testo 6613 6614 6615 6617 N 02 2 m cable le...

Страница 153: ...4 2 Accessories and spare parts 153 Order code Characteristic P 50 Probe length approx 500 mm testo 6612 6613 6614 6615 6617 P 80 Probe length approx 800 mm testo 6612 6613...

Страница 154: ...Testo SE Co KGaA Celsiusstr 2 79822 Titisee Neustadt Germany Tel 49 7653 681 0 E Mail info testo de www testo com 0970 6681 en 05 Vol2...

Отзывы: