![TESTO 6681 Profibus Скачать руководство пользователя страница 106](http://html1.mh-extra.com/html/testo/6681-profibus/6681-profibus_instruction-manual_1092404106.webp)
106
testo 6610
-
2.3 Commissioning
B 2 Duct mounting of testo 6614 heated probe
When mounting the testo 6614 heated probe version, the temperature
probe must also secured at a distance of approximately 10 cm from the
humidity probe. An appropriate assembly tool is supplied with the
testo 6614.
C
Process mounting
•
During installation, ensure that the probe cannot be damaged during
operation.
For the testo 6614 (heated probe version), please use the
cutting ring screw connection, order no. 0400 6193, to
assemble the temperature probe.
Cutting ring screw
connection
Order no 0554 1796
Wall/duct holder
Order no. 0554 6651
testo 6614
approx. 10 cm
distance
Temperature probe
Assembly tool, supplied with
wall/duct holder
Order no. 0554 6651
Direction of flow