110
testo 6610
-
2.4 Maintenance and cleaning
Screw on protection cap by hand, i.e. do not tighten it using a
tool.
2.4.2 Cleaning the instrument and
filter/protection cap
▪
Only clean the instrument carefully with a moist cloth.
▪
Do not use aggressive cleaning agents.
▪
Do not use any solvents.
▪
Unscrew cap or protection cap to clean, clean with compressed air and
replace. Take care not to damage the sensor!
2.4.3 Replacing the sensor
Thanks to the probe concept (digital, plug-in), all probes can be changed in
situ within seconds if necessary, generally without any interruption to the
operation of the machine.
To maintain the extremely high accuracy of the testo 6610
probe, customers are not able to replace the sensor.
Please contact your Testo Service team for this.
Содержание 6681 Profibus
Страница 45: ...testo 6681 1 4 Operation 45 1 4 5 Overview of the testo 6681 user menu...
Страница 46: ...46 testo 6681 1 4 Operation...
Страница 67: ......
Страница 70: ......