P2A software
-
3.3 Using the software
133
The values are reset to the default settings.
5
Click on
[OK]
or
[Cancel]
to close the dialogue.
3.3.4 Adjusting the transmitter
This function is used to adjust an attached instrument. The following
adjustments may be carried out using the software:
•
1-point adjustment (offset)
•
2-point adjustment (upper and lower adjustment point)
•
Analog adjustment (entry via assistant/wizard).
Also see
volume 1, chapter 1.3.3
3.3.4.1 1-point adjustment
1
Expose the reference measuring instrument and the instrument to be
adjusted to the same constant conditions and wait for equalization
period to lapse.
2
Mark the instrument file of the connected instrument.
3
Click on
[Adjusting the transmitter]
.
The
Properties of <Instrument type> <Serial number>
dialogue is
opened with the
Adjusting the transmitter
register.
4
Enter reference value and click on
[Carry out 1-point adjustment]
.
5
Confirm confirmation request.
The adjustment is carried out.