E.2.2 Replacing the probe module
1 Press the key on the top of the probe handle and
remove the probe module.
2 Fit a new probe module and engage it in place.
E.3 Regular care
E.3.1 Condensate trap
The fill level of the condensate trap can be read from the markings on the trap.
A warning message (
red flashing light) is displayed if the level in the condensate trap
reaches 90%.
Emptying the condensate trap
The condensate consists of a weak mix of acids. Avoid contact with the skin. Make
sure that the condensate does not run over the housing.
Condensate in gas path.
Damage tto m
measuring ccells aand fflue ggas ppump.
Do not empty condensate trap while pump is operating.
1 Hold the measuring instrument so that the
condensate outlet points up.
2 Open condensate outlet in condensate trap: Pull
out approx. 5mm or until it will not go any further
( ).
3 Let the condensate run out into a sink ( ).
4 Dab off drops at condensate outlet using a cloth.
5 Close the condensate outlet.
The condensate outlet must be fully closed (mark-
ing) otherwise incorrect measurements due to
inleaking air may result.
Placeholder:
Kondensatfalle_oeffnen.tif
Kondensatfalle_leeren.tif
E. Operation
E.2 Probes/Sensors
19